![Tecnomed Italia Skudo Installation, Operation And Maintenance Manual Download Page 20](http://html.mh-extra.com/html/tecnomed-italia/skudo/skudo_installation-operation-and-maintenance-manual_1072006020.webp)
20
ENGLISH
70°
If necessary, it is possible
to block the SKUDO™ arm in intermedi-
ate positions by acting
on the central knob.
Once in the working position,
engage in the lower opening
the suction conveyor
duly disinfected.
WARNING!
Hold the SKUDO™ head in the centre
and bring it close to the patient's face
DANGER!
Never handle SKUDO™ from the ends of
the arch.Breakage hazard.
Damages attributable to this default are
not covered by the warranty.
SKUDO™ head should be
positioned at a distance of
about 20 cm from the face
so that the practitioner
can work easily in the oral
cavity.SKUDO™ creates an
invisible barrier that blocks
and sucks the aerosols that
pass through it or are in the
vicinity.
To facilitate the work at 9
hours, it is advisable to tilt
the SKUDO™ head by 45°.
In 12 hours, position
SKUDO™ about 10
centimetres from the face
with an angle that allows
agility of action.