Tecnologic TLV 38 Operating Instructions Manual Download Page 4

The instrument allows for measuring calibration, which may be
used to recalibrate the instrument according to application needs,
by using par. 

“OFSt”

 and 

“rot”.

Programming par. “rot”=1,000, in par. “OFSt” it is possible to set a
positive or negative offset that is simply added to the value read by

the probe before visualisation, which remains constant for all the
measurements.

If instead, it is desired that the offset set should not be constant for
all the measurements, it is possible to operate the calibration on  

any two points. 
In this case, in order to decide which values to program on par.

“OFSt” and “rot”, the following formulae must be applied :

“rot” = (D2-D1) / (M2-M1)

“OFSt” = D2 - (“rot” x M2)

where:

M1 =measured value 1
D1 = visualisation value when the instrument measures M1

M2 =measured value 2
D2 = visualisation value when the instrument measures M2

It then follows that the instrument will visualise  :

DV = MV x “rot” + “OFSt”

where:  DV = visualised value 

MV= measured value

Example 1: It is desired that the instrument visualises the value
effectively measured at 20° but that, at 200°, it visualises a value

lower than 10° (190°).    
Therefore :  M1=20 ;  D1=20 ;  M2=200 ;  D2=190

“rot” = (190 - 20) / (200 - 20) = 0,944
“OFSt” = 190 - (0,944 x 200) = 1,2
Example 2: It is desired that the instrument visualises 10° whilst the

value actually measured is 0°, but, at 500° it visualises a 50° higher
value (550°).

Therefore :  M1=0 ;  D1=10 ;  M2=500 ;  D2=550
“rot” = (550 - 10) / (500 - 0) = 1,08
“OFSt” = 550 - (1,08 x 500) = 10
By using par.

 “FiL” 

it is possible to

 

program time constant of the  

software filter for the input value measured, in order to reduce
noise sensitivity (increasing the time of reading).
Using par. 

“diSP”, 

located in the group 

]

PAn”, 

it is possible to set

normal visualization of the display which can be the measure (dEF)
or the alarm threshold AL1, AL2 (AL1, AL2).

4.2 - PEAK VALUES MEMORIZATION AND HOLD FUNCTION

The instrument memorizes the highest and lowest peak
measurement values.  
To visualize such values simply press the 

UP

 key to visualize the

highest peak or the 

DOWN

 key for the lowest peak during normal

operation of the instrument.  
When the instrument is switched off, such values are always re-set.

However, it is also possible to re-set these values if the instrument
is switched on by using the U key that has been suitably

programmed (see par.  U key with function "USrb" = r.Pic).
Again, using the U key it is possible to visualize the difference

between the two peaks on the display (see par. U key  with function
"USrb" = d.Pic).  

Besides the function of the peak values, the instrument also has a
HOLD function through which it is possible to lock the visualization

of the display on the measured value.  
This function can be operated using the U key (see par.  U key with

function "USrb" = Hold ).  
When the HOLD function is actived, the instrument works on the

alarms in operation depending on the memorized measurement.

4.3 - ALARMS OUTPUTS FUNCTIONS (AL1, AL2)

The alarms (AL1, AL2) are depending on the process value and
before to set his functioning it’s necessary to establish to which

output the alarm has to correspond to.
First of all it’s necessary to configure, in the parameters group

]

Out”, 

 the parameters relative to the outputs required as alarm

(“O1F” , “O2F”) programming the parameter relative to the desired
output as follows :

= ALno

 if the alarm output has to be ON when the alarm is active,

while it’s OFF when the alarm is not active

= ALnc

 if the alarm output has to be ON when the alarm is not

active, while it’s OFF when the alarm is active 

= ALni

 if the alarm output has to be ON when the alarm is not

active, while it is OFF when the alarm is active but with reverse led
indication (led ON= alarm OFF).

Note:

 In all the examples that follow is made reference to the alarm

AL1. Naturally the operation of the other alarms results analogous.

Have now access at the group 

]

AL1”,

 and program on par.

“OAL1”

 , to which output the alarm signal has to be sent.

The alarm functioning is instead defined by parameters :

"AL1t "

 - ALARM TYPE

"Ab1"

 - ALARM CONFIGURATION

“AL1”

 - ALARM THRESHOLD 

“AL1L”

 - LOW ALARM THRESHOLD (for band alarm) OR

MINIMUM SET OF AL1 ALARM THRESHOLD (for low or high

alarm)

“AL1H”

 - HIGH ALARM THRESHOLD (for band alarm) OR

MAXIMUM SET OF AL1 ALARM THRESHOLD (for low or high

alarm)

“HAL1” - 

ALARM HYSTERESIS

“AL1d”

 - ALARM ACTIVATION DELAY (in sec.)

"AL1i"

 - ALARM BEHAVIOUR IN THE EVENT OF

MEASUREMENT ERROR

"AL1t" – ALARM TYPE :

  the alarm output can behave in 3

different ways.

LoAb

 = LOW ALARM: The alarm is activated when the process

value goes below the alarm threshold set on parameter "AL1” and

will be deactivated when it goes above the value [AL1+HAL1]. With
this mode is possible to program the minimum and the maximum

set  of  “AL1” by “AL1L” and “AL1H” parameters.

L o A b

o f f

O N

A L 1

A L 1

P V

H A L 1

t i m e

o f f

o f f

O N

O U T

HiAb

 = HIGH ALARM: The alarm is activated when the process

value goes higher than the alarm threshold  set on parameter "AL1"

and will be deactivated when it goes below the value [AL1 - HAL1].
With this mode is possible to program the minimum and the

maximum set  of  “AL1” by “AL1L” and “AL1H” parameters.

H i A b

o f f

O N

A L 1

P V

t i m e

H A L 1

o f f

o f f

O N

O U T

A L 1

LHAb

 = BAND ALARM: The alarm is activated when the process

value goes under the alarm threshold set on parameter "AL1L" or

goes higher than the alarm threshold set on parameter "AL1H" and
will be deactivated when it goes below the value [AL1H - HAL1] or

when it goes above the value [AL1L + HAL1].

L H A b

O N

o f f

A L 1 H

A L 1 L

P V

t i m e

H A L 1

H A L 1

o f f

o f f

O N

O U T

A L 1

TECNOLOGIC spa - TLV 38

 

- OPERATING INSTRUCTIONS - Vr. 01 - ISTR 06612 - PAG. 4

Summary of Contents for TLV 38

Page 1: ...DIAGRAM 3 4 ELECTRICAL CONNECTIONS 3 3 MECHANICAL MOUNTING 3 2 PERMITTED USE 3 1 INFORMATION ON INSTALLATION AND USE 3 FAST PROGRAMMING OF ALARM THRESHOLDS 2 3 PARAMETER PROGRAMMING LEVELS 2 2 PARAMETER PROGRAMMING 2 1 PROGRAMMING 2 FRONT PANEL DESCRIPTION 1 2 GENERAL DESCRIPTION 1 1 INSTRUMENT DESCRIPTION 1 1 INSTRUMENT DESCRIPTION 1 1 GENERAL DESCRIPTION TLV 38 is a digital microprocessor based ...

Page 2: ...oup To exit the programming mode no key should be pressed for approx 20 seconds or keep the UP or DOWN pressed until exit from the programming mode is obtained Longer Hold 2 sec Hold for 2 sec Hold for ATTENTION The instrument is programmed in factory with all the parameters to exception of the Alarm thresholds AL1 AL2 if available programmable in the menù ConF to the purpose to prevent wrong acci...

Page 3: ...n external transformer TCTR or with equivalent features and to use only one transformer for each instrument because there is no insulation between supply and input We recommend that a check should be made that the parameters are those desired and that the application functions correctly before connecting the outputs to the actuators so as to avoid malfunctioning that may cause irregularities in th...

Page 4: ... see par U key with function USrb Hold When the HOLD function is actived the instrument works on the alarms in operation depending on the memorized measurement 4 3 ALARMS OUTPUTS FUNCTIONS AL1 AL2 The alarms AL1 AL2 are depending on the process value and before to set his functioning it s necessary to establish to which output the alarm has to correspond to First of all it s necessary to configure...

Page 5: ...ction Pressing the key for at least 1 sec the display will show the writing 0 Pot for approx 1 sec and then 0 assuming the value measured in that instant as 0 r Pic Pressing the key the highest and lowest peak values are re set r P0P For the instruments with normalized signals input it is possible to set the zero value and contemporarily re set the highest and lowest peak values with this function...

Page 6: ...scription Par 1 000 0 000 2 000 Rotation of the measuring straight line rot 9 0 1999 9999 Measuring Offset OFSt 8 1 0 0FF 20 0 sec Input digital filter FiL 7 C C F Temperature unit of measurement Unit 6 0 Pt1 Ptc ntc 0 1 norm sig 0 3 Number of decimal figures dP 5 Group Out parameters relative to the outputs ALno ALno ALnc ALni OFF Functioning of output 2 see O1F O2F 11 ALno ALno ALnc ALni OFF Fun...

Page 7: ...urther to its expiry please contact our sales department to obtain authorisation for sending the instrument to our company The faulty product must be shipped to TECNOLOGIC with a detailed description of the faults found without any fees or charge for Tecnologic except in the event of alternative agreements 7 TECHNICAL DATA 7 1 ELECTRICAL DATA Power supply 12 VAC VDC 24 VAC VDC 100 240 VAC 10 Frequ...

Page 8: ... 0 20 mA SEnS 0 20 50 0 110 0 C 58 0 230 0 F 50 110 C 58 230 F NTC 103 AT2 SEnS ntc 55 0 150 0 C 67 0 302 0 F 55 150 C 67 302 F PTC KTY81 121 SEnS Ptc 199 9 850 0 C 199 9 999 9 F 200 850 C 328 1562 F Pt100 IEC SEnS Pt1 0 1760 C 32 3200 F tc S SEnS S 0 1370 C 32 2498 F tc K SEnS CrAl 0 1000 C 32 1832 F tc J SEnS J dP 1 2 3 dP 0 INPUT 7 6 INSTRUMENT ORDERING CODE TLV38 a b c d ee f a POWER SUPPLY F ...

Reviews: