Tecnocontrol TS293EO User Manual Download Page 5

IST-2293.EO01.01/C

 

Istruzione / User’s Manual / Manuel d’utilisation 

Pag.5/9

 

 

 

 

TECNOCONTROL S.r.l. Via Miglioli 47 SEGRATE ( MI ) 

 Tel: 02/26 92 28 90   Fax: 02/21 33 734

 

 

Yellow and green Led activates (

0 mA output signal

):

 

this happens when the “

Cartridge Sensor

” is not working. 

First try to perform the procedure of “

 Calibration in Air

” as described in the section

 “

Test and Calibration > Calibration in 

Air

 

then disconnect and connect the unit, finally try to replace a new 

Cartridge Sensor

”. If the condition is not 

change, please replace the unit and/or send it back to the supplier to repair. 

All Led activates 

only if reed Led be activate by Dip-Switch no.1 set to ON

 (

>20 mA output signal

):

 

this happens when the 

Cartridge Sensor

” is not working or gas concentration is higher than F.S. (30%vol. O

2

) If there are not any gas leaks 

and the condition is not change, please replace the unit and/or send it back to the supplier to repair. 

INSTALLATION 

The detector must be accurately installed and testing according to the national dispositions in force on the safety of 
the plants and installation of electric devices in areas with danger of explosion. 

Mounting

: The 

fig. 2

 shows the instrument size. The unit must be positioned vertically with the sensor downwards.  

TS203EO positioning:

 

It should be fixed at 1.60 m from the floor (the 

O

2

 gas is lighter as air). 

Electrical Connection 

(

see fig.2

)

: the maximum distance to install each detector from the Gas Central Unit is show 

in the 

Table 1

. Normally use a tree wire shielded cable. 

The terminals (

+ - S

), are polarized plug-in type, it is necessary to extract them to make the connection. 

Note: Dip-Switch should be set with instrument powered off. Dip-Switch 1 settled ON activates ALARM red 

Led indication (see Table 2). 

Important: 

Once installation is completed,

 

power up the unit, wait about 30 to 60 minutes and then to adjust the 

sensor to the environment, only if it is necessary,

 

carry out the 

Calibration in Air

” 

(see ‘

Test and Calibration

). 

COMPATIBILITY WITH OTHER CENTRAL UNITS: In case of a central unit other then Tecnocontrol, 

please 

verify the max load resistor as shown in 

fig. 4

The max load resistor is 50 ohm with 12Vdc power supply, while is 

400 ohm with 24Vdc power supply. 

We suggest using 24Vdc power supply. 

WARNING 

Average life

: In clean air and in normal working condition the 

sensor's life is

 

about 2 years.

 After this period the 

yellow Led “

FAULT

” flashes every 4 seconds, is necessary replacing the “

Cartridge Sensor

”. 

Periodical testing

: we advise to carry out working tests every 6 or 12 months. as explained in chapter “

Tests and 

Calibration

”. 

Note:

 the detector is not able to detect Oxygen gas occurring outside the room where it is installed. 

Sensor operating limits: 

the electrochemical cell, in any case, should not be exposed to temperatures lower than 

-40°C or higher than 60 °C. The electrochemical cell can be damaged, if it is exposed to concentrations over con-
centration of 30 % O

2

. If this condition occurs, we suggest replacing the “

Cartridge Sensor

”. 

Safety:

 Oxygen sensors contain a small quantity potassium acetate solution which is corrosive. Normally this ma-

terial would only leak out as a result of either mechanical damage or an incorrect instrument use beyond the limits 
of its operational functioning. These sensors also contain small amounts of lead, lead oxide, platinum, silver, car-
bon and antimony. If the user comes into contact with the oxygen sensor contents the affected area should be 
washed with a copious supply of water. 

Cross sensitivity to other gases:

 The electrochemical cell employed has a good resistance towards products 

such as sprays, detergents, ammonia, glues and paints. 

However, in case of products containing substances in 

great quantity, these could interfere with the Sensor and cause false alarms.

 We recommend ventilating the room 

when products like these are used. Volatile organic solvents in great quantity cause false alarms, and then the cell 
needs one or more hours in clean air to recover the normal sensitivity conditions. 

TEST and CALIBRATION 

PAY ATTENTION

: This procedure has to be made with extreme attention and by authorized and trained people; 

because starting this procedure it will increase mA output causing the activation of connected alarm devices. 

Operation Check, Calibration in Air and Fine mA ADJ Output

:

 are different code protected functions. To access 

these functions is necessary to insert the relevant “

Code

" through the keys F1s and F2. To have the key pressure 

recognized, hold pressing it for around a second (

until the Green LED doesn't switch off for a moment

). Then the next key 

can be pressed. In case of error all it takes is waiting around 10 seconds and the sequence is automatically era-
sed. The 

Calibration Check

 is a function code free. 

Calibration Kit, Sample Gas Bottles

 (

for Calibration Check

) please, only using a mixture

 

16% volume Oxygen (O

2

) in nitrogen 

You can use either the disposable cylinder, with adjust valve or

 

the high pressure ones with reduction gear. Is also 

necessary to use 

the calibration kit Tecnocontrol mod. TC011

“INSTRUMENT OPERATION CHECK” (

Check Code

: F2, F2, F1, F1)

this function allows to effect a functional test 

of the detector. After having inserted the "

Code Test

", all Led are switched off. Then Led will switch on in sequence, 

from the yellow up to the red. At the end all the Led will remain lighted for around 5 seconds, then the detector re-
turns at the conditions of normal operation. It is advisable to perform this operation every 12 months according to 
the use. 

Note

: this function is not working if the red Led have already turned on. 

“CALIBRATION IN AIR” (

Calibration Code

: F2, F2, F2, F1, F2, F1)

this function allows completely recalibrate the 

sensor at 20,9%volume in clean air. This procedure when it is launched, will automatically until the end. 

Warning

to guarantee that no errors of elaboration happen, the possibility exists that during the Calibration the yel-

low Led switch off every 8 seconds, in this case interrupt the procedure, switch off and switch on the instrument and 
repeat the Calibration. If condition persists it will be necessary to send the detector to the supplier for the reparation. 

Summary of Contents for TS293EO

Page 1: ...ampo di misura Standard Range Champ de mesure 0 25 0 O2 Limite funzionamento Sensore Sensor limits Limite échelle 30 O2 Vita media in aria pulita Average Life in clean air Vie moyenne en air pur 2 anni years ans Tempo massimo di immagazzinamento Max Storage Time Temps maximum de stockage 3 mesi 3 month 3 mois Tempo di risposta Response Time Temps de réponse T90 30 secondi seconds secondes Ripetibi...

Page 2: ...are Ossigeno ma raggiunge le condizioni di stabilità ottimali dopo circa 15 minuti di funzionamento continuo Funzionamento Normale deve essere acceso il solo Led verde ON Il Led rosso ALARM si accende solo se è stato abilitato con il Dip Switch n 1 posizionato su ON se la concentrazione di gas O2 supera i 22 5 oppure scende sotto il 19 5 I Guasti sono indicati dall accensione del Led giallo FAULT ...

Page 3: ...F1 e F2 Per far sì che la pressione sul pulsante sia riconosciuta tenerlo premu to per circa un secondo finche non si spegne per un attimo il Led verde Dopodichè si può passare al pulsante successivo In caso d errore basta aspettare circa 10 secondi e la sequenza è automaticamente cancellata Kit di Taratura e Bombole con Miscela Gas per la Verifica la miscela da utilizzare è Gas Ossigeno al 15 vol...

Page 4: ...ut of scale or expired NOTES ON THE DETECTED GAS TS293EO has a sensor that can detect Oxygen O2 gas which is colourless odourless a little heavier than air its density as to air is 1 1 Oxygen is not a flammable gas it is a strong oxidant which reacts with reducing and fla mable materials increasing the risk of fire and explosion We all need to breathe the oxygen in air to live Air is made up of se...

Page 5: ...the CartridgeSensor Safety Oxygen sensors contain a small quantity potassium acetate solution which is corrosive Normally this ma terial would only leak out as a result of either mechanical damage or an incorrect instrument use beyond the limits of its operational functioning These sensors also contain small amounts of lead lead oxide platinum silver car bon and antimony If the user comes into con...

Page 6: ...air condition 20 9 v v Then opplying the 16 v v O2 gas insert the TC011 over the sensor holder adjust the sample gas cylinder valve as the flow meter indicates around 0 3 l mins see fig 3 wait for 1 2 minutes verify to read a value of about 143 mV corresponding to 14 3 mA 0 2 output and the Central Unit which is connected to the detector displays about 16 0 2 Then close the gas cylinder and remove...

Page 7: ...de la Cartouche il sera nécessaire d envoyer le détecteur au fournisseur pour réparation INSTALLATION Les détecteurs doivent être installés positionnés et vérifiés en suivant toutes les règles nationales en vigueur pour les installations électriques dans les zones avec dangers d explosions et les normes de sûreté des installations Montage en fig 1 sont indiquées les dimensions Le détecteur doit êt...

Page 8: ... être exécuté en air propre exclusivement milieu sans la présence de gaz toxiques ou autres A l aide des touches exécuter le Code d Etalonnage Attendre que les leds jaune et verte s allument et que la led rouge cli gnote la led rouge reste ensuite en feu fixe pour indiquer que la procédure est en fonction Attendre 3 minutes environ que la led rouge s éteigne au moins durant 2 secondes A ce point d...

Page 9: ...able 2 Tableau 2 S1 SET Dip Switch LED Allarme 1 2 3 4 Alarm LED LED d alarme ON ON ON ON OFF ON OFF OFF OFF ON 190 65 83 105 70 Ø42 107 Fig 1 Dimensioni Size Dimensions Imbocco Femmina Gas Conico UNI 6125 UNI 6125 Female Gas Conical Thread Entrée femelle Gaz Conique UNI 6125 S 12 24V mA 1 2 3 4 ON OFF SET 12 24Vdc Alimentazione Power Supply Alimentation Centrale rivelazione gas Gas Central Unit C...

Reviews: