DDIF051_REV3.doc
20 - 38
PAR 07
ANALYSIS AND RESEARCH OF FAILURES
N° FAULT
CAUSE
INTERVENTION
FEEDING CABLE NOT CONNECTED
TO THE SOCKET OR AT THE BASIS
OF THE MACHINE
CORRECTLY CONNECT
THE FEEDING CABLE
1
DOESN’T PRODUCE
WATER
RIARMO NON ESEGUITO
PREMERE ILTASTO DI
RIARMO
YOU HAVEN’T PUT IN WATER
RIGHT TO THE MAX LECVEL
FILL THE DISTILLER
CORRECTLY
THE GASKET OF THE COVER IS
NOT CORRECTLY PLACED OR
DOESN’T SEAL ENOUGH
PLACE THE GASKET
CORRECTLY
3
DOESN’T PRODUCE
ENOUGH DISTILLED
WATER DURING THE
CYCLE
THE WATER CONTAINER IS NOT
PLACED UNDER THE DRAINING OF
THE DISTILLER
PLACE THE
CONTAINER
CORRECTLY
ACTIVATED COAL FILTERS ARE
NOT PLACED IN THE SPOUT
REPLACE FILTERS
CORRECTLY
4
LEACK OF WATER AND
STEAM FROM THE DRAIN
THE CONTAINER IS NOT
PERFECTLY ALIGNED WITH THE
DISTILLER
ALIGNE THE
CONTAINER