Initial operation
©
BA/ECB/G
B /050 5
9
Allowed ambient tempe5°C - +40°C
Ambient conditions
x
Allowed relative humidity of air up to 30°C: 80%
x
Allowed relative humidity of air up to 40°C: 50%
x
No condensation
Connect the unit to an easily accessible socket as the mains
plug serves as interrupting device!
The Techspan ultrasonic cleaner can be connected within a
range of 100Vac to 240Vac.
Connect the unit to
the mains
Voltage range
100Vac – 240Vac
4.1
Filling of the unit
Fill the cleaning tank with a sufficient quantity of a suitable
cleaning liquid before operation.
Observe filling level
Dry running of the transducer tank can cause damage to the
unit.
Ensure that the cleaning tank is filled to the optimum level
during operation (2/3 filling level).
NOTE
Ensure that the chosen cleaning chemical is suitable for
treatment in an ultrasonic bath and observe the instructions on
dosage and compatibility of the material.
We recommend the use of the cleaning agents listed in section
5.1.
Caution! Flammable products are not allowed for use in an
ultrasonic bath. Observe the instructions given in section 5
(cleaning media).
Restrictions concerning further cleaning media are also
indicated in the instructions given in section 5.
For queries please contact the manufacturer or your supplier.
Freshly mixed cleaning liquids are saturated with air which
lessens the cleaning effect of the ultrasonic activity. By
sounding the liquid over a period of several minutes before the
cleaning process the tiny air bubbles in the liquid are
eliminated.
Suitable cleaning
agents
Prohibited cleaning
agents
Degassing of liquids