background image

- 4 - 

 

Standard calibration process (sender resistance not known): 

Note: Using this technique, it’s necessary to have the sender matching the gauge position being 
calibrated.  For example, the tank needs to be empty when making the gauge read empty, full when 
calibrating full.  It doesn’t need to be calibrated at the extremes, could be ¼ and ¾ for example, but 
best accuracy is achieved when near the end-points.
 

1. Set your sender to a point near one end of the value being measured (we'll start at the top in this 
example, but order is unimportant).

 

2. Set the rotary switch  

Ѳ

 

  to position "1" (High Calibration Point).  The LED in MeterMatch will light 

up. 

3.  Use the Up/Dn buttons to make your gauge read the appropriate setting for the present sender 

output.  Depending on the gauge, the buttons may work backwards, but don't worry about that - just 
make the gauge read the desired value.  Some meters are well-dampened so this process can take a 

little time.  It doesn't hurt to tap lightly on the gauge too.  If you press and hold an  Up/Dn button, the 

gauge signal from MeterMatch will continually move.  It takes about fifteen seconds for the 
MeterMatch to move over its entire range with one of the buttons held down continuously. 

4. Press the  Save button to save this calibration point.  The MeterMatch reads the sender, then 

saves that value, along with the gauge signal, into its internal memory.  The LED on MeterMatch will 
go out momentarily to indicate that the values are being saved, then turn back on. 

Repeat these steps but with the sender set to a position for the other end of measurement, the rotary 

switch  

Ѳ

 

  to "4" (Low Calibration Point), and then use the Up/Dn  buttons until gauge is reading 

appropriately, then press  Save  again.  

Finally, set the MeterMatch rotary switch  

Ѳ

 

 back to "0" (Normal Operation), put the cover back on, 

and installation is complete. 

You can go back and reset any point using the same process, it will simply over-write the data for that 
point stored previously.  If you would like to erase a point (for example a point or limit you don't want 

to use any more):  with no power applied, set the rotary switch  

Ѳ

 

  to the position reflecting the point 

you want to erase, hold down the Save button, then turn power back on.  The LED will blink. 

Alternate calibration process (sender resistance known): 

If you know the resistance of your sender, you can use that to help do calibration in place of making 
the actual sender be in different positions during the calibration.  You will need to know the sender 
range (end-points) in Ohms.   For example, the sender might be 240 

 (Ohms) when the reading is 

low, and 33 

 when the reading is high.  If you have this data, there are two ways that you can use it 

for calibration: 

Summary of Contents for MeterMatch MM-02

Page 1: ...rMatch is wired between your sender sensor and your gauge It reads the resistance of the sender then based on the calibration values controls the current through the gauge to drive it to the desired reading It treats the input and output entirely separately so it doesn t care if the input to the gauge is scaled differently than the sender or even if the sender signal is reversed it can correct for...

Page 2: ...ic process of having the gauge read properly from calibration points near each end of the range is fine for most applications but you can be more sophisticated if you wish For example what if your fuel tank is large at the bottom and small at the top In this case it won t read properly throughout the range MeterMatch provides two additional calibration points between the endpoints So in this case ...

Page 3: ... Pin 3 To gauge output from MeterMatch Pin 2 Ground ensure that it is well grounded Pin 1 12V switched by ignition requires only a few mA Calibration Once wired it is necessary to calibrate your MeterMatch to tailor it to your sender and gauge As shipped default values are in place for switch positions 1 and 4 The rotary switch Ѳ positions are used as follows 0 Normal operation after calibration 1...

Page 4: ...alibration point The MeterMatch reads the sender then saves that value along with the gauge signal into its internal memory The LED on MeterMatch will go out momentarily to indicate that the values are being saved then turn back on Repeat these steps but with the sender set to a position for the other end of measurement the rotary switch Ѳ to 4 Low Calibration Point and then use the Up Dn buttons ...

Page 5: ...o which point you are setting Up Dn buttons to make gauge read the desired setting then Save When the Save button is pressed and the rotary switch Ѳ is not set to 0 MeterMatch will use the resistance value you just entered and then reset itself to back normal operation Calibration Checking and Enhancement Feel free to use a combination of these calibration methods For example you might initially c...

Page 6: ...ary on older slower responding gauges but can come into play with more modern gauges that have less internal dampening This feature can be turned on and off with a power up sequence 1 remove power from the unit 2 set the rotary switch Ѳ to seven 3 press and hold the Save and Dn buttons and finally 4 reapply power to the unit and wait until the power up LED flashes indicating that power up is compl...

Page 7: ...ed in what you make the gauge read with the Up Dn buttons As a result ensure that you have calibrated your gauge accurately rotary switch Ѳ positions 1 4 or your alarm will not be accurate either These points can also be erased using the same procedure as normal calibration points Depending on how your gauge works you may find that the low and high alarm set points are reversed in operation from w...

Page 8: ...ns include a description of the fault to help facilitate repair Risk of damage loss during shipping becomes the customer responsibility once the product is delivered to a common carrier by TechnoVersions MeterMatchTM Specifications Size 1 85 x2 4 x0 93 Weight 1 2 Oz Connections 1x5 screw terminal strip accepting 14 22 AWG wires Input Power Voltage Range 7 Vdc 16 Vdc Sender Input Ohmmeter reading o...

Reviews: