background image

D740 Control Panel 

 
 
 
 

 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Operating Instructions 

 

1.

 

TURN  ON:

  Press  the  ON/OFF  button  once  to  turn  on  and  again  to  turn  off. 

NOTE:

 Hold button down for approximately 1 second to turn unit off.  

 
2.

 

WARM  UP

:  The  detector  automatically  starts  heating  the  sensor.    During  the 

heating  cycle,  the  digital  leak  size  indicator  will  flash  0  and  the  detector  will 
sound a slow “beep”. Warm up is usually less than 20 seconds. 

 
3.

 

READY

: The detector is ready to begin searching for leaks when the flashing 0 

stops and the green sensitivity LED turns on (7g/yr setting). 

The audio “beep” 

increases in frequency and the probe LED begins to blink steadily. 
 

4.

 

SEARCHING:   

When searching for leaks, move the sensor tip along a/c lines 

and fittings no further away than 

3/8” and moving at approximately 3”/second. If 

the detector alarms, make smaller sweeps back and forth until the leak source 
can  be  pin  pointed. 

Note:  The  D740  only  responds  to  changes  in  leak 

concentration. The alarm will re-set automatically if the sensor tip is held 
at  the  leak  source.  If  the  leak  detector  has  been  out  of  use  for  an 
extended period, weeks or months, the following action is recommended. 
Power on the instrument and allow it to come out of warm up. Then run it 
with  the  sensitivity  level  in  the  Hi  (high)  position  for  several  minutes 
before  testing  it  with  the  Leak  Test  Vial.  This  action  will  guarantee  that 
the sensor is fully conditioned for maximum response to refrigerant gas.

 

 

 
 

Digital Leak Size 

Indicator 

Low Battery 

 Indicator 

Sensitivity Level Indicators 

Audio Mute & 

Sensitivity Level Selectors 

Power 

 On/Off 

Summary of Contents for D740

Page 1: ...Model D740 H2 Tracer Gas Leak Detector User Manual Detects H2 N2 5 95 Tracer Gas Design certified by ACSI to meet SAE Standard J2970...

Page 2: ...4a in the HI setting and 7 g yr in the NORM setting while moving the probe at 3 second This complies with SAE Standard J2970 sections 7 2 and 7 3 Leak Testing with Tracer Gas is consistent with the te...

Page 3: ...long a c lines and fittings no further away than 3 8 and moving at approximately 3 second If the detector alarms make smaller sweeps back and forth until the leak source can be pin pointed Note The D7...

Page 4: ...ate of approximately 3 second and as close as possible to the suspected area If looking for an apparently larger leak check first at the 7 g yr Norm setting or the LO setting Once the source of the le...

Page 5: ...in the center of the screw cap to expose the vent hole in the top of the screw cap see fig below 2 Turn on the detector and allow the unit to complete the warm up cycle 3 Place the sensor close to th...

Page 6: ...er battery orientation Sensor Replace Filter Unscrew sensor tip as shown to replace filter Replace filter whenever it becomes visibly dirty or every 2 to 3 months depending on use Replace Sensor Remov...

Page 7: ...itive response Product Specifications Model D740 Name Leak Detector H2 Tracer Gas Sensitivity 4 g yr R 134a equivalent Sensor Life 300 hours Response Time Instantaneous Power Supply 4 AA Alkaline batt...

Page 8: ...with return address phone number and or email address SERVICE CENTER 2651 W 81st Street Hialeah FL33016 WARRANTY POLICY The D740H2 N2 Tracer Gas Leak Detector is warranted to be free of defects in ma...

Reviews: