13
4.
In the start window of the receiver display:
Check the charge status of the batteries the charge
should be no less than 4,0 V). If the battery charge is low,
replace the batteries.
Check accuracy of the
sensor connection. If a
«no sensor» symbol
,
is lit on the display, check
if the sensor is connected
correctly.
Check the received signal type
as «continuous»
(by any of
buttons
▲/▼
)
Set the necessary level of receiver
display backlighting, using
◄/►
buttons
Fig. 5.2
If it necessary, change the
frequency
of the second filter
by changing the number of
harmonic component «
f
HRM
» to
another by buttons or .
№2
5.2 Conduct receiver setup
For tracing of energized
cables please set
frequency 50 Hz.
1.
Switch on the
measurement mode
using the button
5.
Set input signal level using
sensitivity buttons
and
Level (at the bottom of the scale)
must be within 50 ... 90% of the
maximum
2.
To select the necessary
frequency press the frequency
button . symbol will appear
3.
Using buttons
◄/►
, set
the desired frequency in the
filtration zone
for example, 50 Hz
4.
Leave the filter adjustment
mode by pressing the frequency
button . Indicator
will appear in the volume
adjustment zone
6.
Set an acceptable volume
level volume
using
◄/►
buttons
7. Moving along the route,
it is necessary to move the
electromagnetic sensor across
the route in one and the other side
to maintain a maximal level of the
signal.
Attention!
input overload can lead
to the wrong interpretation of data
Summary of Contents for SUCCESS ATP-434N