Water Cut Monitor

Bulletin MNIN003 Issue/Rev 0.0 (5/19)

Installation / Operation Manual

4528 Detector, CX-645 Probe,

CSA Approved 4728 Monitor

MANUAL

Summary of Contents for 4528

Page 1: ...Water Cut Monitor Bulletin MNIN003 Issue Rev 0 0 5 19 Installation Operation Manual 4528 Detector CX 645 Probe CSA Approved 4728 Monitor MANUAL ...

Page 2: ...ation Supervision Start Up and Commissioning Services Available Customer Support Contact Information Customer Service TechnipFMC FMC Technologies Measurement Solutions Inc 1602 Wagner Avenue Erie Pennsylvania 16510 USA P 1 814 898 5000 F 1 814 899 8927 measurement solutions TechnipFMC com TechnipFMC com Literature Library http info smithmeter com literature online_index html ...

Page 3: ... 10 External Powered 10 Typical Applications 12 Typical Tank Flow Diagram 13 Detector Calibration 4528 0 5 VDE Output 14 Detector Calibration 4528 X A with 4 20mA Module Mounted Piggy Back on Detector Card 15 Detector Information Type B and C 17 Installation Hints 17 Troubleshooting 18 Recommended Spare Parts 18 4728 with Explosion proof Enclosure Dimensional Drawing 19 4728 with Weather proof Enc...

Page 4: ...Page 4 MNIN003 Issue Rev 0 0 5 19 Page intentionally left blank ...

Page 5: ...easured by the system Alarm relays and a 4 20 mA output can be used to connect the system to other supervisory or control elements Specifications Input Power 115 VAC 15 50 60 Hz 2 watts 220 VAC Optional 14 28 Volts DC 130 mA S W Range 0 5 0 10 or 0 20 S W Alarm Setpoint 0 100 of S W Range Input Signal 0 5V DC from 4528 Detector Indicators S W Setpoint LCD Merchantable Oil LED Green Excess S W LED ...

Page 6: ...30 sec on 3 sec off Analog Output The analog output is 4 20 milliamperes corresponding to 0 5 0 10 0 20 depending on unit range The 4 20 mA output function is accomplished by the use of a precision integrated circuit The 4 mA point may be adjusted slightly with the zero control on the Model 4728 The 20 mA point is factory preset The Model 4728 utilizes a loop switch to determine whether the 4 20 m...

Page 7: ...1 which is located at the lower right of the P C card The position of this switch determines whether the 4 20 mA output is to be internally powered from the 4728 or externally powered from an outside source Care must be taken to be sure this switch is in its proper position BEFORE an external 4 20 mA loop power source is connected Failure to have the switch in the proper position may cause compone...

Page 8: ... can be accomplished by removing the three screws marked A in the above draw ing and gently lifting the top plate off of the printed circuit board Find the short white or red wire with the clip on it and gently pull the wire straight up Remove the nut from screw B Insert the printed circuit board into the condulet Position the board so the rod sticking up from the probe goes into the round hole on...

Page 9: ...ng the three screws marked A on the drawing Wiring Diagram Note Spade Lugs are not used on CSA Approved Model 4728 Note Typical Interconnecting cable between Detector and Readout Maximum recommended length is 300 feet Installation Operation Manual ...

Page 10: ...al 4 20 mA Loop Warning Loop Switch MUST be up in external position BEFORE connecting external power on 4728 with bottom terminal strip Loop Switch MUST be up before connecting external power on CSA Approved Model 4728 with the two sides and bottom terminal strips ...

Page 11: ... capacitance Nominal Pipe Size Catalog Model Number Stock Number 2 3 4 6 CX 645 200 BGP CX 645 300 BGP CX 645 400 BGP CX 645 600 BGP 81001190 81001195 81001200 81001205 2 3 4 6 CX 645 200 BFP CX 645 300 BFP CX 645 400 BFP CX 645 600 BFP 81001191 81001196 81001201 81001206 2 3 4 6 CX 645 230 BGP CX 645 330 BGP CX 645 430 BGP CX 645 630 BGP 81001192 81001197 81001202 81001207 2 3 4 6 CX 645 260 BGP ...

Page 12: ...he process Percent Water Recording In this application the Indicating Unit provides local reading on an indicating meter and provides an analog output signal to a standard recorder In place of the recorder this output could be connected to a meter computer input or any other DC input device Percent Water Indication and Set Point Control In this application the adjustable alarm may be used to opera...

Page 13: ...Issue Rev 0 0 5 19 MNIN003 Page 13 Installation Operation Manual Typical Tank Flow Diagram ...

Page 14: ... pot is turned 4 Turn the fine zero back counter clockwise CCW approximately 10 turns 5 Turn the coarse zero pot marked ZERO C 25 turns full counter clockwise Listen for clicks 6 Turn the coarse zero CW until the reading is at or near the grindout value Do not tune through zero Always approach the grindout value from the CW direction 7 Next turn the fine zero to get complete agreement of the displ...

Page 15: ...imately 10 turns 9 Turn the coarse zero pot marked ZERO C 25 turns full counter clockwise Listen for clicks 10 Turn the coarse zero CW until the volt meter reads 0 00 to 0 05 Do not turn pot past point where numbers stop getting smaller 11 Next turn the fine zero to get as small a reading possible on volt meter 12 Move Probe Ref switch to Ref position 13 Adjust output span control to make volt met...

Page 16: ... supply a 4 20 mA signal direct to an external readout device It is not designed to be used with a Model 4728 Readout Unit 3 Maximum Loop Resistance is 250 ohms 4 Shielded twisted pair wire is recommended for the 4 20 mA output Ground the shield at one end of the wire only preferably at the Readoutʼ end 5 Starting in mid year 2000 the 4 mA pot adjusting screw will be accessible through a hole just...

Page 17: ...roximately 2 below zero Open Probe Chassis Pickup wire is not connected to the probe Shorted Probe Occurs when electrical short exists in the probe Installation Hints Always install probe up stream of dump valve anytime probe is mounted on oil line coming off treater separator etc Never install downstream as gas will break out to some degree as oil goes from vessel pressure to lower line pressure ...

Page 18: ...I gravity oils reading high est A digital volt meter set on low DC volts range should read less than 25 volts A reading of any voltage indicates the probe insulators are becoming contaminated and should be changed Normally a monitor will operate in a normal manner up to 25 volts but then the readings will become erratic Solution to problem is elimination of coating buildup causes or periodic clean...

Page 19: ...Issue Rev 0 0 5 19 MNIN003 Page 19 Installation Operation Manual 4728 with Explosion proof Enclosure Dimensional Drawing ...

Page 20: ...Page 20 MNIN003 Issue Rev 0 0 5 19 Installation Operation Manual 4728 with Weather proof Enclosure Dimensional Drawing ...

Page 21: ...Issue Rev 0 0 5 19 MNIN003 Page 21 Installation Operation Manual CX 645 Probe Dimensional Drawing ...

Page 22: ...ny P 49 4101 304 0 The specifications contained herein are subject to change without notice and any user of said specifications should verify from the manufacturer that the specifications are currently in effect Otherwise the manufacturer assumes no responsibility for the use of specifications which may have been changed and are no longer in effect Contact Information Field Service Response Center 24 ...