Setting the operating temperature
1. To set the temperature required, press and hold either the up or down button depending on the
direction you need to take the set point.
2. When you have the correct set temperature displayed, release the up or down button and the unit
will start to heat or cool to the set point.
3. Once the process value/actual temperature reaches the set point, allow the block to stabilize for 15
minutes before performing a calibration.
4. The controller is factory pre-set and only the set-temperature should be changed.
THE CONTROLLER MUST NOT BE SET TO CONTROL ABOVE 430°C/806°F OR DAMAGE TO THE UNIT
MAY RESULT.
After use
1. When you have finished calibrating, remember that parts of the unit and the probe may be very hot,
so precautions should be taken to prevent injury.
2.
For safety, set the set temperature and allow the unit to cool to below 50°C/122°F before placing
the unit in a carrying case or shipping carton.
Changing from °C to °F
The controller is preset to display/operate in degrees C; however it may be changed to read in Degrees F. Refer
to section 5-2 of the 2216E Eurotherm Controller manual for instructions on changing from
°C to °F.
Be careful not to change any of the other settings in the configuration menu as performance will be
affected and/or damage to the unit will occur!