background image

 

 

 

Setting the operating temperature 

1.   To set the temperature required, press and hold either the up or down button depending on the  
      direction you need to take the set point. 

2.   When you have the correct set temperature displayed, release the up or down button and the unit     
      will start to heat or cool to the set point.  
3.   Once the process value/actual temperature reaches the set point, allow the block to stabilize for 15  

      minutes before performing a calibration.  
4.   The controller is factory pre-set and only the set-temperature should be changed. 

 

THE CONTROLLER MUST NOT BE SET TO CONTROL ABOVE 430°C/806°F OR DAMAGE TO THE UNIT 

MAY RESULT. 

 

After use 

1.   When you have finished calibrating, remember that parts of the unit and the probe may be very hot,  

      so precautions should be taken to prevent injury. 
2.

 

For safety, set the set temperature and allow the unit to cool to below 50°C/122°F before placing  

the unit in a carrying case or shipping carton.   

 

Changing from °C to °F 

 

The controller is preset to display/operate in degrees C; however it may be changed to read in Degrees F.  Refer 

to section 5-2 of the 2216E Eurotherm Controller manual for instructions on changing from 

°C to °F. 

 

Be careful not to change any of the other settings in the configuration menu as performance will be 
affected and/or damage to the unit will occur! 

 
 

 
 

 

-6- 

Summary of Contents for UCAL400+

Page 1: ...UCAL400 Calibrator OPERATOR S MANUAL Please read all the information in this booklet before using the unit Unit Serial number _________ 4 02 Issue 1 TECHNE is a registered Trademark of Techne Incorpo...

Page 2: ...aluminum insert block as the heat transfer medium The temperature control circuit is built into the unit Features include 10 insert formats available Maximum temperature of 430 C 806 F An independent...

Page 3: ...y 0 015 C 0 025 C Heat up time 20 C to 400 C 12 minutes Cool down 400 C to 100 C 21 minutes Immersion Depth 4 5 114 3mm Fan Cooling Automatic Weight 9 lbs 4 Kg Dimensions H x W x D 8 75 x 8 x 8 inches...

Page 4: ...NDS DO NOT put hot objects on or near combustible objects DO NOT operate the unit close to inflammable liquids or gases DO NOT place any liquid directly in your unit At all times USE COMMON SENSE Oper...

Page 5: ...following value fuse to protect the cable 230V UK 4 AMP The fuse in the unit protects the unit and the operator Note that units marked 230V on the rating plate work at 220V units marked 120V work at...

Page 6: ...ol circuit give good temperature control and uniformity but make sure that there is a close fit of the probes in the block to allow efficient heat transfer 2 Plug the power cable into the socket in th...

Page 7: ...T NOT BE SET TO CONTROL ABOVE 430 C 806 F OR DAMAGE TO THE UNIT MAY RESULT After use 1 When you have finished calibrating remember that parts of the unit and the probe may be very hot so precautions s...

Page 8: ...eaning Before cleaning your unit ALWAYS disconnect it from the power supply and allow it to cool below 50 C Your unit can be cleaned by wiping with a damp soapy cloth Care should be exercised to preve...

Page 9: ...vailable inserts Part Number Insert well Sizes 7022533 insert 5 x 1 4 7032534 insert 1 each 1 8 3 16 1 4 5 16 3 8 7032535 insert 2 x 1 4 2 x 3 8 7032536 insert 2 x 1 4 2 x 1 2 7032537 insert 1 x 1 4 7...

Page 10: ...card This period is from the date of purchase and within this period all defective parts will be replaced free of charge provided that the defect is not the result of misuse accident or negligence Se...