T o u c h g e n e G r a d i e n t O P E R A T O R ’ S M A N U A L
PAGE 36
Interchangeable Blocks
Block removal
Ensure that the unit is turned
off and disconnected from
mains.
The block can be removed and
exchanged for another as
required.
Slide the quick release handles
in the direction shown above
and lift the assembly out of the
unit.
Never lift or carry the block by
one quick release handle,
always use both handles or
support the block underneath.
When refitting the block, or fitting a new one, slide the quick release handles in the opposite direction to
the above arrows to lock the block assembly into the unit. The software will register the change and
calibrate the unit for the new block.
Remember to adjust the options in the Program Defaults Menu for the new application.
If you have fitted the flat block for slide applications you must choose Heated Lid OFF. See the section
headed Program Defaults Menu.