- 10 -
If an error occurs during a measurement, the measurement
is discontinued and a corresponding error code is displayed
(Example: Error No. 2).
Error No. Possible cause(s)
ERR 1
No pulse has been detected.
ERR 2
Unnatural pressure impulses influence the measurement result.
Reason: hand was moved during the Measurement (Artefact).
ERR 3
The inflation takes too long. The cuff is not correctly seated.
ERR 5
The measured readings indicated an unacceptable difference
between systolic and diastolic pressures. Take another reading
following directions carefully. Contact you doctor if you continue
to get unusual readings.
ERR 8
Cuff pressure more than 290mmHg
Other possible malfunctions and their elimination
If problems occur when using the device, the following points should be checked
and if necessary, the corresponding measures are to be taken:
Malfunction
Remedy
The display remains empty
when the instrument is
switched on although the
batteries are in place.
1. Check batteries for correct polarity and if
necessary insert correctly.
2. If the display is unusual, re-insert batteries
or exchange them.
The device frequently
fails to measure the blood
pressure values, or the
values measured are too
low (too high).
1. Check the positioning of the cuff.
2. Measure the blood-pressure again in peace
and quiet under observance of the details
made under point 5.
ERROR MESSAGES/
MALFUNCTIONS
Summary of Contents for KARDIO-TEST KTN-01
Page 2: ......
Page 3: ......
Page 4: ......
Page 5: ...10 0...
Page 6: ......
Page 7: ......
Page 8: ......
Page 9: ......
Page 10: ......
Page 11: ......
Page 12: ......
Page 13: ......
Page 14: ......
Page 15: ......
Page 16: ......
Page 17: ......
Page 18: ......
Page 19: ......
Page 20: ......
Page 21: ......
Page 22: ...1 10 10 0...
Page 23: ......
Page 24: ......