To change the flow rate units, the push button marked with the arrow pointing upwards, 
must be pushed. To change the totalized volume units, the push button marked with the 
arrow pointing left pointing upwards, must be pushed.   
The possible flow rate and totalized volume units are the following: 

Flow rate:

 

There are 9 combinations made from 3 volume units and 3 time units. 
                               

Volume             /          time

  

                                 l (litres)             /     s (second) 
                        m3 (cubic meters)     /     m (minute)  
                             ga (gallons)          /        h (hour) 

Totalized volume:

 

There are 3 possible volume units, l (litres), m3 (cubic meters), ga (gallons) 

Note:

 1ga = 3.785 litres. 

With the working units chosen, by pressing the two push buttons PROG, we go to the loop 
current programming. 

In the first screen we program the flow rate equivalent to 4 mA (lower range). The 
measuring units will be the ones chosen in the previous screen. The push buttons have 
the same function as the pulses per litre programming. 
After that, we program the flow rate equivalent to 20 mA (upper range). 

In a CH420R, if during the programming sequence we receive a HART command that 
must be attended, the local programming will not be valid and all the data previously 
programmed will be lost. The screen will return to the normal working model and the word 
PROG will light on the screen, indicating this event. To remove the word PROG from the 
display, press any of the two PROG push buttons.  

2.2. Serial number indication 

By pressing the three push buttons, the display will show the serial number. To return to 
the usual screen, press any of the push buttons. 

Summary of Contents for CP420R

Page 1: ...Programmable flow rate transmitter CP420R CH420R Instructions Manual R MI CH420R Rev 0 English version...

Page 2: ...There are three push buttons with the following functions 2 1 Programming In all the programming screens the RESET push button exits the screen without saving the data in the memory even if changes in...

Page 3: ...rogramming In the first screen we program the flow rate equivalent to 4 mA lower range The measuring units will be the ones chosen in the previous screen The push buttons have the same function as the...

Page 4: ...in the instrument Example 1 Shortage of volume If we have a flow meter body which specifies i l 1 985 and when we check the volume of a batch we find that instead of having 100 litres as programmed we...

Page 5: ...er of Device Variables 2 Number of Dynamic Variables 1 Mappable Dynamic Variables No Number of Common Practice Commands 13 Number of Device Specific Commands 2 Bits of Additional Device Status 12 Alte...

Page 6: ...digits 7 Size of the digit 8 mm Reset By means of push button 8 3 Flow rate Indication N of digits 5 Size of the digit 5 mm 8 4 General characteristics Ingress protection IP40 Ambient temperature rang...

Page 7: ...imum input frequency is 0 06 Hz 8 6 2 Turbine If the Pulse input is configured for Turbine jumpers in position 1 2 a coil of a magnetic pick up can be connected between terminals 1 2 The input impedan...

Page 8: ...specified in the order the result of service or modification by un authorized persons bad handling or accident This warranty is limited to cover the repair or replacement defective parts which have no...

Reviews: