background image

RECEPTION 

The flowmeter is supplied ready for installation and service.  

The blocking elements that hold the float for transport should be removed before installation.  

Turning the instrument up side down, check that the float moves freely in the tube.    

INSTALLATION 

The instrument must be installed taking into account the following: 

The fluid inlet will be in the bottom of the flowmeter (the one nearest the scale’s minimum value). 

The fluid outlet will be in the top of the flowmeter (the one nearest the scale’s maximum value). 

It is very important that the position of the instrument is completely vertical, given that deviations of 
about 5º– 10º can produce errors of about 10% of the reading. 

Never  open  suddenly  the  regulating  valve  as  this  may  cause  the  float  to  hit  the  glass  tube 
and break it.

3

!

Linea  de lectu ra

Forma AC

F orma  EC G

FLOW RATE READING

The  float  determines  the  flow  rate 
measurement on the scale. 
For  the  different  shapes  of  floats,  the 
readings  must  be  taken  at  the  height 
shown in the drawing at the right. 

CLEANING AND MAINTENANCE

Remove the 4 screws (10) and the transparent protection (9) to gain access inside the body (1). 

Screw up the measuring tube clamp (2) into the body head (1), until the metering tube (7) is free. 

Remove the metering tube (7), taking care that the top and bottom springs (8 and 5) don’t fall out. 

Once the metering tube (7) is removed, withdraw the top and the bottom springs (8 and 5) and the 
float (6). NOTE: The top spring is different from the bottom one and should not be interchanged.  

Cleaning  should  be  done  using  a  soft  brush  (bottle  brush  or  similar)  to  avoid  scratching  the 
measuring tube.  

The float should also be cleaned with a soft brush, never with metallic utensils which could scratch 
it’s surface 

To reassemble the instrument, inspect the gaskets (4) to see if they are in good working condition, 
and if not, change them.  

Put the bottom spring (8) in the measuring tube (7), insert the float and the top spring (5).   

Assemble the measuring tube inside de body. 

Screw down the measuring tube clamp (2) from the body head (1), until the tube is tight and the 
gaskets make a proper seal.

Put the transparent protection (9) in the body (1) and screw in the 4 screws (10).  

AC Shape 

ECG Shape 

Reading Line 