ThermalCapture Fusion – User Manual
4.1.2 Flir Software Installation
The configuration of the Tau core is done by software from FLIR.
You need the FLIR Camera Controller GUI for Tau and Quark:
http://www.flir.com/cvs/cores/view/?id=51880
and the FLIR GUI User Guide:
http://www.flir.com/uploadedFiles/OEM/Products/LWIR-Cameras/Vue/flir_gui_user_guide.pdf
Connect the Mini-USB interface of ThermalCapture Fusion to your PC and follow Chapter 4 of the
FLIR GUI User Guide to connect to the camera.
4.1.3 TAU Configuration
ThermalCapture Fusion gets the images from the digital output of the TAU core. This output is
switched off by default. If the output is not enabled, ThermalCapture Fusion cannot record images.
The digital output must be enabled with the Flir "Camera Controller GUI".
After the FLIR GUI has connected to the TAU core, go to the Digital Video Tab and select CMOS
for XP Bus Output and 14-bit filtered output.
If your TAU core has the “advanced radiometry option”, TLinear should be enabled to have per
10 / 15