Explanation of setting items

General

General

 

Station (Name)

 

 

UR4MD

  In case of

blank, UR4MD is set

 

Institution

(Name)

 

 

 

 

Manufacturer's

Model (Name)

 

 

 

 

SC (Secondary

Capture)

Image Device

ID

 

 

 

 

PDU size

 

 

16384

 

 

Modality

 

 

OT

  In case of blank, OT is set

 

(Storage)

Transfer Retry

Count

 

 

3

 

 

Numbering -

Gender

 

 

Male: 

1

  Female: 

2

  Other: 

3

 

Station (Name)

Enter the station name.

Institution (Name)

Enter the institution name.

Manufacturer’s Model (Name)

Enter the model name.

SC (Secondary Capture) Image Device ID

Enter the secondary capture device ID.

PDU size

Enter the transmission data size.

Modality

Enter the modality name.

(Storage) Transfer Retry Count

Enter the number of times to retry when a storage failure 
occurs.

Numbering - Gender

Enter these values when numbers are used for gender 
with a magnetic card, for example.

 

o

The unit will compare genders entered using a 
remote control or card reader, for example, to these 
setting values, and convert them to M, F, and O.

DICOM Storage

DICOM Storage

 

Mode

 

 

OFF   ON(Still only) 

 

IP address

 

 

192

  . 

168

  . 

0

  . 

150

 

 

Port

 

 

104

 

 

AE title

 

 

STORE_SCP

 

 

Calling AE title

 

 

UR4MD_SCU

 

Mode

Select ON (Still only) to use the storage function.

IP address

Enter the DICOM storage server IP address.

Port

Enter the DICOM storage server port number.

AE title

Enter the DICOM storage server AE title.

Calling AE title

Enter the UR-4MD AE title.

General

11

Summary of Contents for UR-4MD

Page 1: ...e Z Thank you for choosing TEAC Read this manual carefully to get the best performance from this unit After reading this manual keep it in a safe place for future reference UR 4MDDICOM Digital Video R...

Page 2: ...ault address conflicts with any existing address a usable address can be manually assigned General Contents General 2 Settings 3 Getting started 3 How to obtain an IP address 3 Obtain an IP address au...

Page 3: ...Ethernet Cable connecting to an Ethernet Port in the back Computer UR 4MD Network Server How to obtain an IP address This section describes how to obtain an IP address If the network supports DHCP the...

Page 4: ...3 Press the ENTER button to determine Automatic DHCP Please wait dialog is briefly shown 4 Move the cursor to Current IP Address and press the ENTER button to show IP Address Take note of the IP addre...

Page 5: ...ith the existing address connect a USB keyboard to the USB 3 port back of the unit to input a different IP address When finished press the ENTER button to determine the default or newly input address...

Page 6: ...n the computer 2 Enter the UR 4MD IP address that was obtained in the previous process followed by 10080 in the browser address bar Example http 192 168 0 121 10080 A dialog box shows up asking for a...

Page 7: ...Activation When the browser shows the UR 4MD screen perform the following steps 1 Click DICOM Service 2 The browser shows the Activation Key input dialog General Continued on the next page e 7...

Page 8: ...for how to check the MAC address o o Make sure that the UR 4MD is in the idle screen date and time displayed before entering the activation key Otherwise an error message will appear as shown below E...

Page 9: ...on the browser screen These cannot be set with the settings screen of the unit itself UR4MD Digital Video Recorder About Device status Set up ECHO test DICOM service DICOM settings Administration Cur...

Page 10: ...n case of blank UR4MD is set Institution Name Manufacturer s Model Name 2 After you have finished entering the setting items click the Save changes button at the bottom of the screen ATTENTION Press t...

Page 11: ...nter the modality name Storage Transfer Retry Count Enter the number of times to retry when a storage failure occurs Numbering Gender Enter these values when numbers are used for gender with a magneti...

Page 12: ...cit VR Little Endian IP address Enter the MWM server IP address Port Enter the MWM server port number AE title Enter the MWM server AE title Calling AE title Enter the UR 4MD AE title Requested Date S...

Page 13: ...settings Administration Current settings Log le Date Time ECHO test DICOM C ECHO C ECHO test Manual setting IP Port Called AE Calling AE OK C ECHO test Storage OK C ECHO test MWM OK Press the OK butt...

Page 14: ...ity Worklist Management function By enabling the Modality Worklist Management func tion the Worklist can be acquired and shown on the UR 4MD s front display The patient information selected from the d...

Reviews: