34
5-3-3. Zero Balancing
With no load on the sensor, press the ENT button.
During zero balancing, a pop-up appears to show that it is in
progress. When zero balancing completes, the setting screen
for the next procedure appears.
o
If a calibration error appears, conduct countermeasures
according to the error message.
5-3-4. Rated capacity (load calibration)
Set the actual load value and conduct load calibration.
With an actual load applied to the sensor, press the ENT
button.
After measuring the sensor output, confirm the rated capacity.
o
If a calibration error appears, conduct countermeasures
according to the error message.
ATTENTION
The Rated Capacity is set to the D/A Full Scale value.
NOTE
The decimal point position set here will be used as the indica-
tor value decimal point position.
5-3-5. Select EU
Select the unit that corresponds to the indicator value.
Options: N, kN, kPa, MPa, g, kg, sht, ton, mN·m, N·m, kN·m, none
5-3-6. Max. Disp. Value
Set the highest displayed value (page 36).
5-3-7. Cal. Value Lock
To prevent unintended changes to the calibration value, Cal.
Value Lock should usually be set to ON.
5. Calibration