36
5-4-4. Select EU
Options: N, kN, kPa, MPa, g, kg, sht, ton, mN·m, N·m, kN·m, none
Select the unit that corresponds to the indicator value.
o
If a sensor with TEDS built in is connected, the read unit
will be set.
5-4-5. Cal. Value Lock
To prevent unintended changes to the calibration value, Cal.
Value Lock should usually be set to ON.
ATTENTION
If Cal. Value Lock is ON, automatic calibration will not occur
when the unit is turned on.
5-5. Sensor value memory
Use this to select the sensor value memory.
Set setting value memory settings using System settings 1
(page 55) Equivalent Input Calibration, Actual Load
Calibration and TEDS calibration
5-6. Select Min. Grid
Set the minimum digital change of the indicator value.
Options: 1, 2, 5, 10
5-7. Max. Disp. Value
Set the highest displayed value.
The default value is set at 110% of the Rated Capacity.
If this value is exceeded, “±FULL: Plus Value Over (Over Max.
Disp. Value)” appears in a pop-up message.
ATTENTION
The input value will be checked with a value converted to
match the input signal.
5. Calibration