TRUBLUE USER’S MANUA
L
Water Level, Conductivity, and Barometric Pressure Data Loggers
TE CONNECTIVITY SENSORS
///
TRUBLUE USER’S MANUA
L
08/2016
Page 24
Appendix A - TruBlue Connection Diagrams - Continued
THE 853-
XXXXX
INTERFACE CABLE
The 853 Interface Cable
Interface Cable is intended for use with any data logger as a long term
deployment solution. The assembly incorporates an innovative vent filter that provides lifetime moisture protection
for vented gage water level data loggers. A GORE-TEX plug prevents water droplets from entering the filter while
allowing air to pass freely to the internal silica gel desiccant. The filter, housed in corrosion-resistant grade 316
stainless steel, provides surge protection, and features a compact Hirose electrical connector.
Hirose to Backshell
design allows for the user to leave a vented data logger deployed in the field, while still maintaining the ability to
extract data.
Note:
T
he 853
–
XXXXX
Interface Cable requires the 850-00857 USB to Hirose communication cable for
operation.
NOTE:
When deploying a data logger with an Interface
Cable Assembly, it is very important to make sure the
electrical connector is covered with the supplied cap.
GORE-TEX FILTER
HIROSE CONNECTOR WITH CAP