7. Sealed Construction 

Because of the TruBlue data logger’s sealed design (which improves durability by eliminating moisture 

intrusion), the unit’s batteries cannot be removed, replaced or recharged. The two 3

-volt lithium ion 

batteries that are built into the unit will last more than 5 years when logging 1 reading every 15 minutes. 
For applications where data must be recorded at a higher frequency we offer additional Auxiliary Battery 
Packs and Backshells as accessories. The Auxiliary Battery Pack provides an additional 7.2V Lithium 
battery that is field installable and replaceable. To accommodate the additional battery pack, Auxiliary 
battery pack backshells are offered in both 316 Stainless Steel and Titanium.  

8. Internal Clock Synchronization 

When received from the factory, your TruBlue data logger’s internal calendar/clock will be set to Eastern 

Standard Time. The internal clock can be quickly and easily synchronized to your local clock and time 
zone using TruWare. Please refe

r to the TruWare User’s Manual for details.

 

9. General Field Installation 

The TruBlue is calibrated for operation in a vertical, upright position with the cable extending up. Although 
it may be operated in any orientation, anything other than this vertical position will result in a slight zero 
offset in the measured data. If the data logger will be operated in a known, non-vertical orientation this 
zero offset error may be easily calibrated out. 

10. Cable Protection 

 Cable damage is one of the most common causes of vented data logger failure. Here are a few 

precautions you should take in order to extend the life of your TruBlue data logger: 

• When working with your data logger cable, make sure the cable does not drag over sharp edges. This 

could nick or cut the cable outer jacket. A cut in the outer jacket could allow water into the data logger 
electronics causing permanent damage. 

• Our polyurethane jacketed cables are quite flexible; however, care needs to be taken to avoid crimping 

the vent tube inside the cable when bending the cable to suit your installation. To avoid this, do not bend 
your cable tighter than a radius of 3 inches. 

• If you use a compression fitting or strain relief to secure the cable as it enters a junction box, be careful 

not to over-tighten the fitting to the extent that it damages the cable or crushes the internal vent tube.  

WARNING. 

Crimping the vent tube with tight bends or crushing it within the cable can block the 

reference pressure port of the water level data logger. This will prevent the reference from 
registering atmospheric or temperature-related pressure changes, resulting in measurement 
offset errors

 

 

Summary of Contents for TruBlue 255 LEVEL

Page 1: ...CONNECTIVITY SENSORS TRUBLUE USER S MANUAL 08 2016 Page 1 TruBlue User s Manual FOR WATER LEVEL CONDUCTIVITY TEMPERATURE AND BAROMETRIC PRESSURE DATA LOGGERS 255 LEVEL 275 BARO 285 CT 555 LEVEL 585 C...

Page 2: ...1 9 General Field Installation 11 10 Cable Protection 11 11 Conductivity 12 Types of Conductivity Measurement 12 Actual Conductivity 12 Specific Conductivity 12 Calibrating the Conductivity Sensor 12...

Page 3: ...HNICAL SUPPORT While this manual provides extensive guidance for everything from setup through maintenance we understand that there s nothing like getting an expert on the phone to provide quick answe...

Page 4: ...L Level Temperature High Accuracy Yes Vented or Absolute 585 CTD Conductivity Level Temp Yes Vented or Absolute KEY FEATURES User friendly TruWare software included free of charge Fully sealed design...

Page 5: ...s the software program designed specifically to manage TruBlue data loggers and the data they collect Used together a TruBlue data logger and TruWare are ideal for monitoring groundwater and surface w...

Page 6: ...schedules for connected data loggers Extract data from a previously initiated test View save and export test data and real time readings View data as a table or dynamic self updating graph displaying...

Page 7: ...E 852 STAINLESS STEEL SUSPENSION WIRE PART NUMBER 852 XXXXX Is a low cost suspension solution intended for use only with absolute transducers Wire is available in configurable lengths from 0 999 feet...

Page 8: ...le in the other near the backshell While applying slight pressure to push the two parts together slowly rotate the data logger assembly in the backshell until the keys align When that happens the hous...

Page 9: ...provides lifetime moisture protection for vented gage water level data loggers A GORE TEX plug prevents water droplets from entering the filter while allowing air to pass freely to the internal silic...

Page 10: ...ld It features the Hirose connector to quickly link to the vent filter connector and unique USB RS 232 conversion connector at the opposite end This same cable also provides external power to the data...

Page 11: ...able extending up Although it may be operated in any orientation anything other than this vertical position will result in a slight zero offset in the measured data If the data logger will be operated...

Page 12: ...CTIVITY UNITS microsiemens per centimeter S cm default millisiemens per centimeter mS cm A means of expressing what the actual conductivity of a solution would be at a standard reference temperature 2...

Page 13: ...a Meter View window displaying a conductivity reading A With the TruBlue inserted in the calibration solution hold the mouse pointer over the conductivity reading and right click to get to the Conduc...

Page 14: ...been chemically attacked eroded or physically altered A slow degradation of the sensor is hard to spot until it s too late Cleaning should be done with a combination of soaks and rinses in a solution...

Page 15: ...ransients can travel distances of a mile or more 3 TRANSDUCER RESPONSE TO PRESSURE LEVEL INPUT CHANGES BECOMES SLUGGISH This is usually a sign that the sensing end of the transducer has become fouled...

Page 16: ...aning solution Growth can be reduced by utilizing the 860 00825 Anti Fouling Guard to protect the sensor You can also coat the transducer with marine grease This may be the most effective and inexpens...

Page 17: ...766 1500 or toll free at 1 800 745 8008 during normal working hours Eastern Standard Time Be prepared to supply the following information when requesting the RMA Data logger part number Serial number...

Page 18: ...iagrams THE 850 00857 USB TO BACKSHELL COMMUNICATION CABLE The 850 00857 USB to Backshell Communication Cable is universal to vented gage and absolute data loggers This NON SUBMERSIBLE communications...

Page 19: ...cludes built in conversion to RS 232C and USB as well as a DC input connector The 3 foot coiled cable can remain connected to an RS 232C modem port and communicate with a PDA handheld via USB Connect...

Page 20: ...n Diagrams Continued THE 850 00856 USB TO HIROSE COMMUNICATION CABLE The 850 00856 USB to Hirose Communication Cable is the newest and most streamlined cable offering This cable provides connection fr...

Page 21: ...data is stored in the on board memory Polled data is sent to a network capable auxiliary logger or telemetry system that capture the live data and transmits from a remote location These devices often...

Page 22: ...Continued THE 852 XXXXX STAINLESS STEEL SUSPENSION WIRE The 852 suspension wire assembly is intended for use with absolute transducers to suspend the unit This assembly includes a stainless steel or t...

Page 23: ...ny Non Vented data logger as a long term deployment solution This Direct Hirose to Backshell design allows for the user to leave a non vented data logger deployed in the field while still maintaining...

Page 24: ...droplets from entering the filter while allowing air to pass freely to the internal silica gel desiccant The filter housed in corrosion resistant grade 316 stainless steel provides surge protection a...

Page 25: ...dix A TruBlue Connection Diagrams Continued THE 860 00825 ANTIFOULING SCREEN Extend TruBlue conductivity data logger deployments in high bio fouling environments with a TruBlue Anti fouling Screen mad...

Page 26: ...sing out of the sale resale use or misuse of the product Users of TE Connectivity products should make their own evaluation to determine the suitability of each such product for the specific applicati...

Reviews: