USER NOTICE M210
TE CONNECTIVITY SENSORS
/// USER NOTICE M210_EN 03/2018
Page 3
3. Display
3.1 SETTING THE ZERO
•
Use the front panel potentiometer (bottom left hand side) to adjust the analogue output to 0
Vdc with the sign flashing.
3.2 SETTING THE DISPLAY GAIN
•
The gain is set using the calibration function and the gain potentiometer. The gain
potentiometer is hidden behind the front panel on the bottom left hand side, above the zero
potentiometer.
•
The calibration function simulates a pressure, or force, as determined during calibration.
•
To simulate the pressure (or force) press the calibration pushbutton (see section
II
of the
"ADJUSTMENT" section if no pushbutton is present).
•
Use the gain potentiometer to obtain the calibration value on the display.
•
Release the pushbutton, check the zero and refine the adjustments if necessary.
The M210 is ready to function