background image

System 3

6-7

PZ2 PreAmp

last LED will flash red. TDT recommends charging the battery before this flashing 

low-voltage indicator comes on. While charging, the Level LEDs will flash green.

Charging

 

the

 

Batteries

Operate the preamplifier with the charging cable disconnected. Connecting the PZ2 

charger will simultaneously charge both batteries. TDT recommends putting the three-

position switch in the 

OFF 

(middle) position while charging the PZ2. 

Charging

 

Indicators

When powered on, the PZ2 battery status LEDs are also used for each battery to 

indicate which battery, if any, is charging. These LEDs are found next to the Level 

LEDs by the headings -A- and -B-. A green indicator denotes the battery bank is 

fully charged while a red indicator designates the battery is currently charging. When 

the device is in operation (charger is not connected) the -A- and -B- LEDs are 

not lit.

An external battery pack is also available to provide longer battery life for extended 

recording sessions. See “PZ-BAT External Battery Pack for the PZ Amplifiers” on 

page 6-23. 

PZ2

 

Technical

 

Specifications

Status

Description

8 Green

Fully Charged

1 Green, 7 Unlit

Low Voltage

1 Flashing Red

Low Voltage - Charge 

Immediately!

8 Green Flashing

Charging in Progress

Status

Description

Red

Charging

Green

Fully Charged

Unlit

Operation Mode (charger not connected)

A/D

Up to 256 channels, 18-bit hybrid

Maximum Voltage In

+/- 10 mV

Frequency Response

3 dB: 0.35 Hz – 7.5 kHz

6 dB: 0.2 Hz – 8.5 kHz

Anti-Aliasing Filter

4th order Lowpass (24 dB per octave)

S/N (typical)

73 dB

Summary of Contents for PZ2

Page 1: ...channel models Note When sampling at a rate of 50 kHz only the first 128 amplifier channels will be available System Hardware All PZ2 channels are organized into groups of 16 channel banks with each b...

Page 2: ...hile collecting data Powering ON To turn the preamplifier on move the three position battery switch located on the front panel of the PZ2 to either the Bat A or Bat B position Powering OFF To turn the...

Page 3: ...o indicate a clip warning Green Activity Red Clip Warning Clip Warning When the input to a channel is greater than 3dB from the preamplifier s maximum voltage input the LED for the corresponding chann...

Page 4: ...eads to remove any potential RF noise that might travel through the cable For best results position the ferrite bead close to the source of the RF noise Battery Overview The PZ2 preamplifier features...

Page 5: ...indicator denotes the battery bank is fully charged while a red indicator designates the battery is currently charging When the device is in operation charger is not connected the A and B LEDs are not...

Page 6: ...on all headstages should be connected together to form a single common ground See Headstage Connection Guide on page 6 99 Distortion typical 1 A D Sample Rate Up to 48828 125 Hz Sample Delay Dependen...

Page 7: ...me Description Pin Name Description 1 A1 Analog Input Channels 14 V Positive Voltage 1 5V 2 A2 15 GND Ground 3 A3 16 GND 4 A4 17 V Negative Voltage 1 5V 5 Ref Reference 18 HSD Headstage Detect 6 HSD H...

Page 8: ...6 10 System 3 PZ2 PreAmp...