er Artifact During MRI Scans
We recommend the following setup:
Place the PZ5 / PZA Subject Interface ampli
fi
er outside the MRI room in the control room or
component room. You can place the ampli
fi
er on a cart in the control room just outside the
penetration panel. All cables pass through either a waveguide on the penetration panel or
through a DB25 connector on the panel.
Turn off the wi
fi
and the front display screen on the PZ5 / PZA Subject Interface ampli
fi
er.
See
using a phantom with no RF
fi
lter. This is your baseline signal.
If there is signi
fi
cant artifact connect the RF
fi
lter to the back of the PZ5 / PZA Subject
Interface ampli
fi
er and the GND on the RF
fi
lter to either:
a. The MRI GND. This is the GND found on the gradient power supply. This GND is
fl
oating
relative to earth and will remove most of the artifact.
b. Room or Shield GND. This is the GND found in the penetration panel, which is connected
to the shielded room around the MRI. This shield usually goes to truth earth GND. This will
remove some but not all the artifact.
1.
2.
3.
4.
fMRI User Guide | 23