7. Mounting

Carefully insert the device into the panel cut-out. Insert the mounting bracket from
the back and push towards the panel until the device sits tight. Make sure the
mounting   bracket   is   snapped   into   the   side   of   the   housing.   To   ensure   IP65
protection (dust and water jet) when mounting in a front panel, use optional gasket
(separately available).

8. Specifications

DPM72-PP

Supply

12 - 24 V AC/DC ±10% (50/60 Hz ±10%)

or via USB, galvanically isolated

Power consumption

Max. 1.2 W

Display

LCD graphic display 192 × 160 pixels

Measuring range voltage

±50/60 mV AC/DC shunt, 10 - 500 Hz

Accuracy voltage

±0.5 % true RMS

Internal resistance voltage

20 kΩ

Measuring range ampere

20 mA AC/DC (max. 100 mA), 10 - 500 Hz

Accuracy ampere

±0.5 % true RMS

Internal resistance ampere

6.2 Ω

Measuring value update

5 Hz (32 kHz sampling rate)

Recording of measurement

36 seconds to 14 days,

180 internal memory locations

Alarm outputs

2 optocoupler outputs max. 30 V DC, 50 mA

Permissible voltage between 
supply and measuring inputs

< 50 V AC, < 75 V DC

Operating temperature

0 °C to +50 °C

Storage temperature

-20 °C to +70 °C

Front panel

72 mm × 72 mm

Panel cut-out

68 mm × 68 mm

Mounting depth

51.8 mm 

Protection front

IP65 from the front

9. Cleaning

Observe the safety instructions before cleaning the device. Clean the device with a
dry lint-free soft cloth. Do not use solvents.

10. Contact Information

TDE Instruments GmbH, Gewerbestraße 8, D-71144 Steinenbronn
phone: +49 7157-20801
E-mail: [email protected]
Internet: www.tde-instruments.de, www.digalox.com

----- 

8

 -----

Summary of Contents for Digalox DPM72-PP

Page 1: ...device hazardous voltages must not be connected to the device The device must not be used as the only protective device or protective shutdown Read instruction manual carefully before operating the d...

Page 2: ...ion display style pointer tachometer bar graph and more splash image thresholds for alarm output hysteresis and more Recorded measurement values can be read and a continuous transmission of the actual...

Page 3: ...Digalox Load Shunt U 50 V max 200 mV Ilow Ihigh Uhigh Ulow L N Legend Current source e g active sensor or sensor with voltage supply WARNING The device may only be operated in one of the connection o...

Page 4: ...display value or the one that is selected by the jumpers J4 J6 or the DIP switches 11 12 respectively see table display value Scaling via DIP switches If the scaling to represent the correct display v...

Page 5: ...lues repeat steps 3 8 until all required display values are configured The time base does not need to be set again as it is used for all display values 10 In operation the display values can be change...

Page 6: ...2 3 4 5 6 7 8 9 1 0 1 1 1 2 Table display value factory preset can be changed by software Display value Display Scale Manual mode Save mode 60 mV DC scaled shunt Ampere DC Upper scale caption ON 1 2 3...

Page 7: ...sec When first opening and closing J1 the current time base is displayed For each subsequent opening and closing the time base changes to the next setting In order to save the setting permanently DIP...

Page 8: ...mpere 20 mA AC DC max 100 mA 10 500 Hz Accuracy ampere 0 5 true RMS Internal resistance ampere 6 2 Measuring value update 5 Hz 32 kHz sampling rate Recording of measurement 36 seconds to 14 days 180 i...

Reviews: