3. Description

The measurement  types  volt AC/DC, ampere  AC/DC, frequency and 5 A current
transformer are supported. Scale endpoint of the indicator as well as scaling of the
analogue signal and current transformer measurement can be adjusted freely via
the DIP switch, alternatively there is an option for automatic adaption of the scale
endpoint. The unit can be switched between volt and ampere measurement via an
external switch. Minimum and maximum values are recorded and can be displayed
optionally using an external switch. Measured values are recorded over a time
span of 36 seconds  up to 14 days. The time base as well as the display of the
measuring history can be activated by an external switch. The measured values
remain stored as long as the device is supplied with voltage.
The   following   parameters   can   be   adjusted   using   the  configuration  software
“Digalox

®

  Manager”:   scale   endpoint,   scale   caption,   display   style   (pointer,

tachometer,   bar   graph,   and   more),   splash   image,   thresholds   for   alarm   output,
hysteresis,   and   more.   Recorded   measurement   values   can   be   read   and   a
continuous transmission of the actual measuring value can be enabled. Using the
software, values can be viewed and exported as CSV.

Front

1

Unit or free text

2

Threshold values

3

Upper scale caption

4

Graphical measurement display

5

Digital measurement value

6

Lower scale caption

Back

1

2 optocoupler switch outputs

2

DIP switch for setting upper scale 
caption and display value (mode)

3

Supply voltage input

4

Jumper pins J1 – J6 for 
activating/switching of graphical 
historic data display, min/max 
display, auto-scaling and display 
value (mode)

5

USB Micro-B interface

6

Measurement inputs

----- 

2

 -----

Summary of Contents for Digalox DPM72-AVP

Page 1: ...evice hazardous voltages must not be connected to the device The device must not be used as the only protective device or protective shutdown Read instruction manual carefully before operating the dev...

Page 2: ...main stored as long as the device is supplied with voltage The following parameters can be adjusted using the configuration software Digalox Manager scale endpoint scale caption display style pointer...

Page 3: ...ansformer voltage direct Load Current Transformer Ilow Ihigh Uhigh Ulow max 10 A FF 10 A L N max 500 V Current on high side 50 V max Ilow Ihigh L N max 10 A max 50 V Uhigh Ulow FF 10 A Load Current on...

Page 4: ...splayed only until the power supply of the DPM72 is interrupted In further operation the device always shows the first display value or the one that is selected by the jumpers J4 J6 or the DIP switche...

Page 5: ...lues repeat steps 3 8 until all required display values are configured The time base does not need to be set again as it is used for all display values 10 In operation the display values can be change...

Page 6: ...5 6 7 8 9 1 0 1 1 1 2 J1 J6 ON 1 2 3 4 5 6 7 8 9 1 0 1 1 1 2 J1 J6 A AC direct Ampere AC ON 1 2 3 4 5 6 7 8 9 1 0 1 1 1 2 J1 J6 ON 1 2 3 4 5 6 7 8 9 1 0 1 1 1 2 J1 J6 V DC direct Volt DC ON 1 2 3 4 5...

Page 7: ...2 sec When first opening and closing J1 the current time base is displayed For each subsequent opening and closing the time base changes to the next setting In order to save the setting permanently D...

Page 8: ...e 2 6 M Measuring range ampere 10 A AC DC and 5 A AC for current transformer 10 500 Hz Accuracy ampere 1 true RMS Internal resistance voltage 5 m Measuring range frequency 10 1000 Hz Accuracy frequenc...

Reviews: