TDE Instruments Digalox DPM72-AV2 Instruction Manual Download Page 3

5. Product overview

3

5

4

2

1

2

6

Front

1 Unit or free text
2 Threshold values
3 Upper scale caption
4 Graphical measurement display
5 Digital measurement value
6 Lower scale caption

4

1

5

3

2

Back

1 Measurement inputs
2 Measuring resistor (hot!)
3 Jumper pins J1–J6

for activating/switching of 
graphical historic data display, 
min/max display, auto-scaling, 
measurement mode

4 Supply voltage input
5 Interface (optional)

6. Mounting

Carefully   insert   the   device   into   the
panel   cut-out.   Insert   the   mounting
bracket   from   the   back   and   push
towards the panel until the device sits
tight. Make sure the mounting bracket is
snapped   into   the   side   of   the   housing.
To   ensure   IP65   protection   (dust   and
water   jet)   when   mounting   in   a   front
panel,   use   optional  gasket  (separately
available).

-----

 3 / 8 

-----

Press mounting 

bracket

Panel thickness: max. 6 mm

Summary of Contents for Digalox DPM72-AV2

Page 1: ...is not isolated from the measurement circuit At the maximum current strength 10 A 500 mW of waste heat is generated at the current sense resistor The tip of the current sense resistor becomes very hot...

Page 2: ...an be displayed and reset using an external switch Measured values are recorded over a time span of 36 seconds up to 14 days The time base as well as the display of the measuring history can be switch...

Page 3: ...ric data display min max display auto scaling measurement mode 4 Supply voltage input 5 Interface optional 6 Mounting Carefully insert the device into the panel cut out Insert the mounting bracket fro...

Page 4: ...rrent and voltage direct Load max 10 A FF 10 A Digalox U U I U I N L max 300 V Current via current transformer Load Current transformer FF Digalox U U I U I N L max 10 A Current via current transforme...

Page 5: ...T J1 J6 To configure the upper scale caption connect J3 within 10 seconds after restart and disconnect it again The device switches to programming mode Prog appears in the display followed by the uppe...

Page 6: ...ged by alternately opening and closing J1 interval 2 sec When first opening and closing J1 the current time base is displayed For each subsequent opening and closing the time base changes to the next...

Page 7: ...0 500 Hz Accuracy ampere 1 true RMS Internal resistance ampere 5 m Measuring range frequency 10 1000 Hz Accuracy frequency 0 1 Hz Measuring value update 5 Hz 32 kHz sampling rate Recording of measurem...

Page 8: ...nstruments Digalox DPM72 EX2 USB interface 15 Contact information TDE Instruments GmbH Gewerbestra e 8 D 71144 Steinenbronn Phone 49 7157 20801 E mail info tde instruments de Internet www tde instrume...