background image

Configuration of the upper scale caption

The upper scale caption is binary coded using DIP switches 1-9. Possible values are 
1 to 500. Switch 1 corresponds to 256, switch 2 corresponds to 128, switch 3 corres-
ponds to 64, etc., switch 9 corresponds to 1. For configuration, proceed as follows:

1. Connect supply voltage.
2. Set DIP switches 1-9 to OFF position.
3. Set DIP switch 10 to ON position.
4. Start with 1st switch.
5. Set switch to ON position.
6. If the displayed value is greater than the desired value, set the switch back to 

OFF position.

7. If the displayed value is less than the desired value, leave the switch in ON posi-

tion and move on to the next switch.

8. Repeat steps 5 to 7 until the desired value is displayed.

----- 

5

 -----

Summary of Contents for Digalox DPM72-AV

Page 1: ...ement inputs of the device can carry life threaten ing voltages WARNING When working on the device hazardous voltages must not be connected to the device The terminals J1 J6 are not isolated from the...

Page 2: ...ernal switch The measured values remain stored as long as the device is supplied with voltage 2 4 3 1 Front 1 Measurement unit 2 Upper scale caption 3 Graphical measurement display 4 Digital measureme...

Page 3: ...rrent via current transformer and voltage direct Digalox Load Current transformer Ilow Ihigh Uhigh Ulow max 10 A FF L N max 500 V Current via current transformer on high side and voltage direct Digalo...

Page 4: ...easurement modes via an external switch Save mode The required measurement modes are configured and saved sequentially 1 Set DIP switch 10 11 and 12 to OFF position 2 Connect supply voltage 3 Set the...

Page 5: ...iguration proceed as follows 1 Connect supply voltage 2 Set DIP switches 1 9 to OFF position 3 Set DIP switch 10 to ON position 4 Start with 1st switch 5 Set switch to ON position 6 If the displayed v...

Page 6: ...4 5 6 7 8 9 1 0 1 1 1 2 J1 J6 ON 1 2 3 4 5 6 7 8 9 1 0 1 1 1 2 J1 J6 A AC direct Ampere AC ON 1 2 3 4 5 6 7 8 9 1 0 1 1 1 2 J1 J6 ON 1 2 3 4 5 6 7 8 9 1 0 1 1 1 2 J1 J6 V DC direct Volt DC ON 1 2 3 4...

Page 7: ...e setting permanently DIP switch 10 must be switched from ON to OFF position J2 Min Max display The display shows the maximum and min imum values recorded since the last reset The values can be reset...

Page 8: ...requency 10 1000 Hz Accuracy frequency 0 1 Hz Measuring value update 5 Hz 32 kHz sampling rate Recording of measurement 36 seconds to 14 days 180 internal memory locations Operating temperature 0 C to...

Reviews: