2
Copyright 2017 Thermoplastic Comfort Systems, Inc.
Operating Instructions: TCS Manual Injector
www.tcsdentalinc.com / 866-426-2970
I092614 Rev B
Effective Date: 05/15/2017
tcs
1. Select a workbench that is firm and very sturdy.
2. Use 4 screws to secure the press to the workbench.
FIG 1
3. Screw the three handles tightly on the press, do not leave loose as
this may damage the handles.
FIG 2
4. Select a place near the injector to place a furnace.
FIG 3
Note:
The furnace should be close enough to the press to allow for easy
transfer of the molten resin from furnace to press.
FIG 3
Note:
Refer to TCS Digital Furnace Operating Manual for instructions on
how to use the furnace in conjunction with this injector.
5. Use the flask guide to center the flask for injection.
FIG 4
Place the
flask vertically with the injection sprue opening facing up directly
under the piston.
FIG 4a
How to position the flask if the flask does not have a guide:
1. Unscrew the flask guide reference from the injector unit. Keep the
screw in case you want to use it at a later date.
2. Visually center the flask with the injection sprue opening facing up
directly under the piston.
3. Position the cylinder vertically over flask, hold in place with one hand
and turn handles of press so the piston goes into the cylinder. This
will be your guide to ensure the flask is centered. Be extra careful to
center the flask when not using the guide.
FIG 4b
tcs
tcs
WATLOW
FUSE
POWER
Digital Furnace
R
SD31
1
SET
2
%
I
o
tcs
FIG 1
FIG 2
FIG 3
FIG 4a
FIG 4
sprue
opening
flask
guide
flask
guide
tcs
FIG 4b
3.
Getting Started
General Safety Hazard:
Refer to the instructions for details on the specific hazard.
Caution:
Hot surface Warning
Caution:
Electrical Shock Hazard
Technical Symbol:
All operations marked with this symbol are to be performed by qualified maintenance
personnel only.
Your safety and the safety of others are very important. We have provided many important safety messages in this manual. Please read these
messages carefully. A safety message alerts you to potential hazards that could hurt you or others. Each safety message is associated with a safety
alert symbol. The definition of these symbols are described below:
2.
Unpacking and Safety
Ventilation Clearance:
AREA
MINIMUM REQUIRED CLEARANCE
Back of the unit
10 cm / 4 inches
Sides of the unit
20 cm / 8 inches
Above
30 cm / 12 inches
Below
N/A