![tbs electronics ATP1-120 Instruction Manual Download Page 21](http://html1.mh-extra.com/html/tbs-electronics/atp1-120/atp1-120_instruction-manual_3820869021.webp)
14
CHAPTER
5
–
INSTALLATION
AND
OPERATIONAL
QUALIFICATION
This procedure describes the steps for installation and testing of the basic
mechanics of the unit and should be done on all new units or units returning from
factory service.
1. Position the unit on a flat level surface away from heat (radiators, stoves,
direct sunlight, etc.) and moisture (sinks, drains, etc.).
2. Connect the unit to a power source. The main power supply connection must
provide (depending on the model):
110 VAC +/- 10%; 60 Hz; 10A;
grounded
220 VAC +/- 10%; 50 Hz; 6A;
grounded
In addition to the ATP
anti-blackout feature, an uninterruptible power
supply (UPS) can be utilized to provide power for short power outages
and some protection against power fluctuations, line noise and power
spikes. Use of a UPS is highly recommended for sites which
experience “brown-outs” or other power fluctuations.
The use of adapters to plug in more than one instrument to the same
outlet (power strip) does not provide a reliable power supply. It may
produce overheating and create a potential fire hazard. ECC electrical
instrument directives forbid such practices.
3. Turn the power on.
4. Select the preferred Language on the initial screen.
5. From the
Main
screen select
Service/ Edit Setup
. Set the elevation to the
appropriate value for the location of the laboratory. (See Chapter 12 for more
details.)
This is particularly important for higher elevation location. Failure to set
the elevation could result in short fill and pressure errors.
6. Choose your Alarm settings.
7. From the
Main
screen select
Service/ Pressure Test.
Perform a pressure
test to confirm the operation of the pumps and sensors associated with the
pressure system.