± 0.5 kV, ± 1 kV
differential
mode line-line
± 0.5 kV, ± 1 kV
differential
mode line-line
Surge
IEC 61000-4-5
Mains power quality
should be that of a
typical commercial or
hospital environment.
0 % UT
(100 % dip in UT )
for 0.5 cycle at 0°,
45°, 90°, 135°,180°,
225°, 270°, and
315°
0 % UT
(100 % dip in UT )
for 1 cycle at 0°
70 % UT
(30 % dip in UT )
for 25/30 cycles at
0°
0 % UT
(100 % dip in UT )
for 250/300 cycle at
0°
0 % UT
(100 % dip in UT )
for 0.5 cycle at 0°,
45°, 90°,
135°,180°, 225°,
270°, and 315°
0 % UT
(100 % dip in UT )
for 1 cycle at 0°
70 % UT
(30 % dip in UT )
for 25/30 cycles
at 0°
0 % UT
(100 % dip in UT )
for 250/300 cycle
at 0°
Voltage dips,
short interrup-
tions and volta-
ge variations
on power supply
input lines
IEC 61000-4-11
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user of
the models TT-CL016
product name requires
continued operation
during power mains
interruptions, it is
recommended that the
models TT-CL016 be
powered from an
uninterruptible power
supply or a battery.
30 A/m, 50/60Hz
30 A/m, 50/60Hz
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Power frequency
magnetic fields should
be at levels characteristic
of a typical location in a
typical commercial or
hospital environment.
NOTE:
UT is the a. c. mains voltage prior to application of the test level.
EN
EN
03
05
07
09
11
13
15
17
19
21
23
25
27
29
31
33
35
37
39
41
43
45
47
49
51
53
55
57
59
61
63
65
67
69
71
73
75
77
79
81
83
85
87
89
91
93
95
97
99
02
04
06
08
10
12
14
16
18
20
22
24
26
28
30
32
34
36
38
40
42
44
46
48
50
52
54
56
58
60
62
64
66
68
70
72
74
76
78
80
82
84
86
88
90
92
94
96
98