3
|
Technical Parameters
Power Supply:
110V, 60Hz
Fuses:
15 AMP x 2
Power Consumption:
1000VA
Operating Temperature Range:
15
Cooling System:
DC 12/24 V Compressor
Refrigerant:
R134A
Dimensions (LxDxH):
26.5” x 22.75” x 16.5”
Weight:
77lbs
Paraffin Reservoir:
6L
Thermal Chambers:
Approx. 1.5L
Cold Plate:
Approx. 80 Cassettes
Functional
Unit
Temperature
Range
Heated
Separately
Separate
Temperature Control
Warm-Up Time
Prior to Operation
Paraffin Reservoir
0°C to 85°C
+
+
2 hrs
Paraffin Dispenser
0°C to 85°C
+
+
15-20 min
Thermal Chambers
0°C to 85°C
+
+
15-20 min
Hot Plate
0°C to 85°C
+
+
15-20 min
Cold Plate
0°C to -10°C
+
+
30 min
Forceps Holders
0°C to 85°C
—
—
15-20 min
Working Environment
• Operate in an ambient temperature range of 5
°
-40
°
C.
• Humidity should be less than 80%.
• Atmospheric Pressure: 860 hpa – 1060 hpa.
• All heated components include overheat protection.
Temperature Control
4 Tanner Scientific® TN1500 - Embedding Center
Summary of Contents for TN1500