1.3.2
Loading Unit Outputs
Each loading unit has two or three output taps. The highest currents are available on the
lowest voltage taps.
Loading
unit
Open
Circuit
Voltage
Continuous
current
Intermittent current
5 min on/
15 min off
1 min on/
15 min off
40s on/
15 min off
NLU5000
2.3V
1500A
3000A
4500A
5000A
4.6V
750A
1500A
2250A
2500A
9.2V
375A
750A
1125A
1250A
NLU2000
3.5V
600A
1200A
1800A
2000A
6.9V
300A
600A
900A
1000A
13.8V
150A
300A
450A
500A
NLU200
60V
40A
80A
160A
200A
120V
20A
40A
80A
100A
Each output is rated for a continuous current and also for higher intermittent currents. The
maximum ‘on’ time for intermittent currents is enforced by the unit, and the output will be
switched off if the rating is exceeded. The display will then show the following message until
the green ‘off’ button is pressed to clear the trip condition:
Output Tripped
Press Green to Reset
NOTE:
If the metering cable is not connected between the control unit and the external loading unit,
the ammeter will not function and the output will not switch on when the output
ON
push-
button is pressed. The metering cable between the control and external loading unit acts as
an interlock circuit.
The control unit automatically selects the correct metering ranges and protection for the
loading unit connected.