21
5.
OVERALL PERFORMANCE SPECIFICATION
5.1
Insulation resistance at 1000V DC
Not less than 10 megohms between mains input and frame.
5.2
Applied voltage test
2kV RMS for 1 minute between mains input and frame
5.3
Accuracy of instruments
KV meter ±1.5% at full scale
mA meter ±1.5% at full scale
5.4
H V Transformer
2.5kV RMS for 1 minute between primary and earth
2.5kV RMS for 1 minute between earthy end of secondary and earth
5.5
Complete Equipment
Over voltage test
Equipment run for 5 minutes at 115% of normal output voltage
Flash over-test
6 flash-overs direct to ground at 100% of normal output voltage
Summary of Contents for KV30-40
Page 2: ...2 ...
Page 4: ...4 ...
Page 6: ...6 ...
Page 13: ...13 2 OPERATION 2 1 Connections 2 2 Front Panel Functions 2 3 Method of Operation ...
Page 22: ...22 ...
Page 24: ...24 ...
Page 26: ...26 ...
Page 28: ...28 Drawings Required A2 000308 latest issue A2 000542 latest issue ...