Basic Function
- TR-51S/52S
44
Adjustment Settings: from the [Communication] Menu
Here it is possible to record data after applying an adjustment formula to
the measurement.
1
.
Set the Data Logger so that communication can be carried out with the
Computer.
2
.
In the [Communication] Menu, click [Adjustment Settings] to open the
settings window.
[Communication] Menu
[Send Settings]
Button
[Initialize]
Button
Check one, and enter the value
[History] Column
3
.
Place a checkmark next to either [
1
Point Adjustment] or [
2
Point
Adjustment] and enter the [Before] and [After] values.
-For more about guidelines for adjustment, see p.
30
.
4
.
By clicking the [Send Settings] button, a confirmation message will
appear.
5
.
Click the [OK] button to transmit the settings to the Data Logger.
[Default] Button
This will return to settings with no adjustment applied.
[History] Column
The latest
4
adjustment history records will be displayed. The records show the date
and time that adjustment settings were made, the adjustment point(s), and the values
for "Before" and "After" adjustment.