Product Specifications
RTR-53A
Measurement Channels
1 Temp and 1 Humidity Channel
Measurement Item
Temperature and Humidity
External Temp. / Humidity
Sensor
0 to 55
℃・
10 to 95%RH
Measurement Display
Resolution
0.1
℃・
1%RH
Measurement Accuracy
Avg.
±
0.3
℃
±
5%RH {at 25
℃
and 50%RH}
Recording Interval
1,2,5,10,15,20,30 seconds,
1,2,5,10,15,20,30,60, minutes,
Total of 15 choices
Recording Capacity
8000 readings
×
2 Channel
Recording Mode
Endless
(Overwrite oldest data when capacity is full)
・
One-time
(Stop reading when capacity is full)
LCD Display Items
Current Readings
・
Recording Settings
・
Battery Life Warning
・
Over Measurement Range Warning
・
Unit
Power
Lithium Battery (LS14250(SAFT)
×
1 or AC Adaptor (sold separately)
Battery Life
Approx. 6 months / Approx. 2.5years with RTR-05B1
(Battery life depending on measurement environment, recording
interval and battery performance)
Wireless Method
ETSI EN 300-220
Transmission Distance
Up to 100 meters (May vary with conditions)
Interface
Wireless Communication
・
Optical Communication
Communication Speed
When downloading data (Wireless) Approx. 2,000 readings
per minute [Collection of a full unit of data=Approx. 420
seconds (optical communication=Approx. 160 seconds)]
Water Resistance
IP64 (rated for use in daily life)
※
1
Dimensions
H62 x W47 x D19mm (excluding protrusions / antenna length
20mm) / with Large Capacity Battery Pack: D50mm
Weight
Approx. 56g. (including 1 lithium battery) / with Large Capacity
Battery Pack : Approx. 109g
Unit Temp. Resistance
−
40
℃
to 80
℃
※
2
Standard Sensor / Input Cable
TR-3310
(length: Approx. 1m)
×
1
Accessories Included:
Lithium Battery (LS14250 (SAFT))
×
1
・
Tube
×
1, Strap
×
1
・
User's Manual (Warranty)
×
1
※
1 Note: The Water resistance rating with the sensor or input cable connected is IP64; not for
continued immersion.
※
2
Radio commnication can not be operated at below -30
℃
degrees or less.
User's Manual
Thank you for purchasing this product. Carefully read
and fully understand these instructions before using
this unit.
Wireless Temperature Data Logger
●
Please read this manual carefully before using the product.
●
All rights of this User’s Manual belong to T&D Corporation. It is prohibited to use,
duplicate and/or arrange a part or whole of this User’s Manual without the
permission of T&D Corporation.
●
Please follow the safety precautions carefully. We cannot guarantee nor are we
responsible for safety if this product is used in any manner other than was
intended.
●
T&D Corporation accepts no responsibility for any malfunction of and/or trouble
with this product or with your computer that is caused by the improper handling of
this product and will deem such trouble or malfunction as falling outside the
conditions for free repair of the attached warranty.
●
T&D Corporation accepts no responsibility for any result or effects from using this
User’s Manual.
●
Figures and illustrations in this manual may be slightly simplified and may differ
from the actual product.
●
On screen messages, figures or illustrations in this manual may vary slightly or
be simplified from the actual messages and product.
●
We sincerely hope that the contents of this manual are true and complete. If you
find any information to have been omitted, or if the information within is confusing
or mistaken please, contact your retailer or T&D Corporation.
●
Microsoft
®
, Windows
®
and WindowsNT
®
are registered trademarks of Microsoft
Corporation USA and are binding in the USA and all other countries. Company
names and product names are trademarks or registered trademarks of each
company.
●
This User’s Manual cannot be reissued, so please keep it in a safe place.
●
Please carefully read this User’s Manual and Warranty.
◆
Notices about this User’s Manual
◆
Safety Precautions and Instructions
※
Please carefully observe the following safety measures when using our
product.
To prevent any loss or damage to our customers, other people and/or property, and to ensure the proper use of our products we ask that before using
our product you carefully read, understand and follow the safety rules and precautions for our products as outlined below.
【
Explanation of Warning Symbols
】
DANGER
These entries are actions that absolutely under no
circumstance should be taken. The taking of such an action
may cause serious personal physical damage or death.
CAUTION
These entries are actions that if taken may lead to physical
injury or damage to persons or things.
【
Explanation of Picture Symbols
】
This symbol denotes an important warning or caution. Inside or near the
symbol will appear another symbol giving details.
(EX:
stands for ELECTROCUTION)
This symbol denotes a forbidden action. Inside or near the symbol will
appear another symbol giving details.
(EX:
stands for DO NOT TAKE APART)
This symbol denotes an action that you must take. Inside or near the
symbol will appear another symbol giving details.
(EX:
stands for TAKE PLUG OUT OF SOCKET)
DANGER
The RTR-53A is a devices to measure temperature humidity. Do
not use this unit for any purpose other than temperature humidity
measurement.
If any smoke or strange smells are emitted from the unit, immediately
cease using it.
Continued use may cause fire, electrocution or damage.
If water or a foreign object enters the case, immediately cease using
it.
Please be careful when using in overly hot or cold environments,
touching the units may cause burns or frostbite.
Store all batteries, sensors and Thermo Recorder units out of the
reach of children. It is dangerous to swallow batteries.
Do not take apart, repair or modify the main unit.
It may cause fire, electrocution or damage. Ask the shop where you
purchased the products or T&D Corporation to carry out any repairs.
Do not use any batteries other than those that are recommended. It
may cause fire or damage.
CAUTION
This is printed on recycled paper.
Wireless Regulations
Do not drop or expose the unit to strong impact.
Do not use or store the unit in places such as listed below:
It may cause electrocution, fire or damage to the unit or to your
computer.
●
Areas exposed to direct sunlight
●
Areas exposed to water or high-pressure water flow.
●
Areas exposed to organic solvents and corrosive gas.
●
Areas exposed to strong magnetic fields
●
Areas exposed to static electricity.
●
Areas exposed to fire or overheating.
●
Areas exposed to excessive dust or smoke.
This product has been designed for private or industrial use only.
It is not for use in situations where strict safety precautions are
necessary such as in connection with medical equipment whether
directly or indirectly.
We are not responsible for any malfunction or trouble caused by the
use of our product or by any problem caused by the malfunction of
our unit. Please be fully aware of this before using our product.
Battery life depends on the measurement environment,
communication frequency, recording interval and battery quality.
Battery terminals may provide insufficient contact due to age or
vibration. Please be careful not to lose data due to insufficient
contact.
Pay attention to water leakage or foreign objects entering into the
unit case as in the following cases.
●
The case was closed with dust, hair, etc., on the rubber packing or in the
groove for the packing.
●
The rubber packing becomes damaged. (In this case, please purchase
the optional maintenance set.)
●
The unit suffered from significant temperature change while wet.
(Especially if the temperature change is from high to low.)
Remove batteries from any unit that is not to be used for a long period
of time.
Batteries left in a unit not being used for a long time may leak and cause a
malfunction.
About the TR-3310, the standard temperature and humidity sensor for
RTR-53A.
●
Use the sensor only within the measurable temperature and humidity
range (Temperature: 0 to 50
℃
, Humidity: 10 to 95
%
RH)
●
The Temperature/Humidity sensor cable cannot be extended.
●
When the sensor is not used, put it in the attached plastic bag with a
drying agent and keep it in cool, dark place at 5 to 25
℃
and 30
%
RH or
less.
●
The service life of humidity sensors can vary greatly depending on
operating environment.Periodic calibration may be required.
Please use the latest version of our software which can be
downloaded free of charge from our Website.
Radio, EMC and Safety Regulations
・
RTR-5 Series complies with technical specifications
required under EN 301 489-3 (with battery and AC
Adaptor), EN 300-220-3 and EN 60950:2000.
・
Allowed to use in : A.B.D.DK.F.I.P.S.SW.UK.N.NL.
CH.FIN.AUS.NZ
Do not put your fingers or foreign matter into the sensor connection.
Avoid using the lithium batteries, LS 14250 for long periods at
temperatures over 60
℃
.
The battery life may be significantly decreased.
Cracks may develop in the unit casing if it comes into contact with
hazardous substances such as oil products.
When using the unit in such an environment, protect the unit by
placing it inside a polyethylene bag.
© Copyright T&D Corporation. All rights reserved. 2011.08 16504350001 5th Edition
Shimadachi 817-1, Matsumoto, Nagano 390-0852 Japan
Tel:+81-263-40-0131 Fax:+81-263-40-3152
E-mail: [email protected]
Homepage: http://www.tandd.com/