TP-40BA (Rev 1.0)
- 24 -
TALON TECH CO. LTD.
6-4 Notice
6-4-1 Degas Function
Degas Overtime : Degas Pump’s Overtime counts. If the real pumping is more than the Degas
setting count and the empty sensor still senses as empty, the error occurs. At this time, you set the
overtime counts. If you want 10 times of pumping, input ‘10’.
Notice) If the data sets “0”, even if Degas Pump keeps pumping, no alarm message.
6-4-2 Dispense Cycle
During the system or the manual dispense, the pump doesn’t save Recipe changes and setting
changes. At this time, ‘Busy’ window is shown normally.
6-4-3 Pump ID Setting
The basic ID is ‘11’. If pump & touch pad is set in the first time, connect pump & touch pad as 1:1
not to double ID. ex) Pump1 : [11], Pump2 : [12], Pump3 : [13], Pump4 : [14], Pump5 : [21].
Otherwise, pump cannot be searched or although pump is searched, the setting data are
overlapped or Data Error / System Error occur. Before setting Pump ID, check that ID is valid or
not.
6-4-4 Recipe Setting
In case Recipe is not set properly, there is “Write Recipe Error” window.
But, this window is shown in only case each total dispense volume is not same as reload volume.
Other cases are applied as normal. So, be careful for “Dispense Time” setting.