3.Edit the PCR program
Touch a 'PCR Program Editor' from Home screen.
Change the User
Accounts to click this part.
Show a standard program on the screen.
Touch an 'Load File' icon on the bottom shelf.
Select a objective program file to load and
click'
✔
'button.
Touch a column of temperature, reaction time
and number of cycles. Appear the numeric
keypad screen to enter numeral of each column.
Touch an 'Add Step' icon, Add new event
after positive blue column or the end of
profile or start of profile that you select.
Check 'Insert after xx Within Cycling',
Insert to select a 'Hold' or 'Pause' step after
positive blue column within cycling
Pause event hold the temperature at the last
set point and display time is 'Infinite'.
Hold event is a simple hold at given
temperature for a set time.
Regulate the time rates reaching the set temperature
to check 'Use Custom Setting' on 'Limit the Ramp Rate
to the Temperature'.