TFM VT-222 Vortex Flowmeter User Manual
point value;
Enter the password 02080 to enter:
Standard gas
pressure
(gauge
pressure)
Standard Pre.
Gas gauge
pressure
direct
input
Unit: Current pressure unit
Standard gas
temperature
Standard Temp.
gas
temperature
direct
input
Unit: current temperature unit.
Flow
correction
factor
Flow correction
factor
Flow
correction
factor
direct
input
Global K-Factor to adjust the indicated flow value
over the entire range. Value can be from 0.1 to 10
Unit correction
factor
Unit correction
factor
Unit
correction
factor
direct
input
Unit K-Factor to adjust the SPCL units to allow
using SPECIAL units. This should be a U32/10000
so we can cover mScf/h and mSCF/d Value can be
from 0.1 to 10
Enter the password 02086 to enter the test mode:
Traffic test
Flow Test
Traffic test
direct
input
Unit: HZ (1-5000)
Note 1: Calculate the corresponding instantaneous
flow rate, current value, and pulse (frequency) value
according to the input sensor frequency value, and
output it.
These values are also read through RS485.
The cumulative flow advances just like the real mode
for validation.
Note 2: The test mode will only be executed after a
new value must be entered, and the LCD backlight
will flash at this time.
Frequency
testCurrent test
Frequency Test
Frequency
test
direct
input
Unit: HZ (10-5000)
Note 1: This function only outputs the specified
frequency value. The instantaneous flow rate value
and current value at this time are not affected. The
cumulative flow remains the same.
Note 2: The test mode will only be executed after a
new value must be entered, and the LCD backlight
will flash at this time.
Current test
Loop Test
Current test
direct
input
Unit: mA (3.9-21)
Note 1: This function only outputs the specified
current value. The instantaneous flow rate value and
pulse output value at this time are not affected. The
cumulative flow remains the same.
Note 2: The test mode will only be executed after a
new value must be entered, and the LCD backlight
will flash to indicate in test mode.
TFM VT-222 Vortex Flowmeter Quick Manual Rev 10f page 20 of 30