![TA DISCOVERY DSC2500 Getting Started Manual Download Page 34](http://html1.mh-extra.com/html/ta/discovery-dsc2500/discovery-dsc2500_getting-started-manual_3112362034.webp)
Discovery DSC™ Getting Started Guide
Page 33
Installing the DSC Cooling Accessories
The DSC operates in conjunction with one of several cooling accessories. Your choice of cooling accesso-
ries depends on the temperature range required for your experiments. This section provides the information
needed to install the FACS and Quench Coolers. For proper operation, the correct cooler type must be
selected through the instrument control software. See the online help for details.
NOTE
: For information on installing the RCS90 and/or RCS40, consult the appropriate manual or the
online help.
Installing the Finned Air Cooling System (FACS)
Follow the instructions below to install the FACS onto the Discovery DSC.
1
Ensure that the cell is at a temperature where contact with the cell will not cause a burn.
2
If an Autosampler is present, it must be moved away from the cell by pressing Lid Open/Close on the
touch screen or on the controller
3
Carefully lift up and pull the top cover off of the Discovery DSC.
Figure 11
Remove cover.
WARNING: When installed, the cell ring restricts airflow to the FACS. As a result, the cell ring
may reach an unsafe handling temperature, posing a burn risk to the Discovery DSC operator.
AVERTISSEMENT: Une fois installé, l'anneau de la cellule restreint l'écoulement de l'air vers le
FACS. Par conséquent, l'anneau de la cellule peut atteindre une température dangereuse pour la
manipulation, présentant ainsi un risque de brûlure pour l'opérateur de Discovery DSC.
WARNING: Do not place your hands in the path of the Autosampler when it is in motion. Physi-
cal injury may occur.
AVERTISSEMENT: Ne placez pas vos doigts sur le chemin de l'échantillonneur automatique
lorsqu'il est en mouvement. Des blessures physiques peuvent se produire.
Pull up on tab to remove cover