Page 50
AR-G2/AR 2000ex/AR 1500ex Getting Started Guide
Maintenance and Repair
CAUTION: Adjustment, replacement of parts, maintenance and repair should be carried out by
trained and skilled TA personnel only. The instrument should be disconnected from the mains
before removal of the cover.
Moving the Instrument
Please follow these recommendations when you move or lift the instrument and its accessories:
•
Always remove the temperature control module from the rheometer before attempting to move it.
Details on how to do this can be found in TRIOS Online Help.
•
When moving the rheometer, the bearing clamp should always be in place, ensuring that the bearing
cannot be moved.
1
Insert the draw rod into the top of the rheometer.
Figure 21
Inserting the draw rod.
WARNING: The cover should only be removed by authorized personnel. Once the
cover has been removed, live parts are accessible. Both live and neutral supplies
are fused and therefore a failure of a single fuse could still leave some parts live.
The instrument contains capacitors that may remain charged even after being dis-
connected from the supply.