48
Troubleshooting
Malfunction
Cause
Solution
The control module’s
battery indicator does
not light up when con-
nected.
The battery fuse is defec-
tive or improperly mount-
ed.
Change the defective fuse /
clean the contacts and insert
it properly.
The power plug is not con-
necting with the power
supply.
Inspect the connectors.
When the wheelchair
is turned on, the bat-
tery indicator blinks
and sounds an acoustic
alarm.
One or both of the drive
motors are set on push
mode.
Move the drive/push mode
levers on both sides to drive
position.
The connector on of the
drive motors is not proper-
ly connected.
Inspect the connectors.
Electronics issue.
Must be repaired at an author-
ised repair shop. Push mode
Drive/push mode lever in
drive position.
Other malfunctions
Refer to ‘Troubleshooting’ in
the control module operating
manual.
Lights are not working.
Defective LED bulb
Let an authorised repair shop
perform the repair/replace-
ment.
Defective control or light-
ing module
Let an authorised repair shop
perform the repair/replace-
ment.
MDR,
not
final
version