Technical specifications

1. 

Basic specifications of the device:

measurement on steel, galvanized steel and aluminium;

• 

measurement resolution: 1 or 10µm;

• 

measurement range: from 0µm to 1100µm;

• 

measurement memory (also after turning the device off): 100 units;

• 

zeroing function;

• 

LCD backlight;

• 

intuitive operation with 8-positions menu;

• 

automatic shut-off after longer idleness;

• 

measuring sensor diameter: 15 mm;

• 

powered by 9V alkaline battery (e.g. 6LR61) or corresponding accu;

• 

power consumption: ~50 mA..

• 

Getting started

2. 

Coating thickness gauge TW-6  is designed to measure the thickness of a steel, galvanized 
steel or aluminium sheet paint coating. The measurement resolution equals 10µm. The device is 
equipped with an internal EEPROM memory for 100 measurements which enables to check the 
results after the measurement is made.  The memory doesn’t get deleted after turning off the 
device, although it is possible to delete it from the menu. The device has a LCD backlight which 
makes measuring in places with insufficient light (e.g. garage) much easier.

Before You start to use the device, please place a new alkaline battery or a fully charged accu in 
the back side of the device’s housing. In order to do it please take off the battery flap and place 
correctly the battery or accu by connecting it to the clips on the cable. 

WARNING! 

T

he

 

devIce

 

Is

 

pReseT

 

To

 

GAlvANIzed

 

sTeel

 

ANd

 

AlumINIum

  (z

N

/A

l

)!  s

hould

 

The

 

meAsuRemeNT

 

be

 

TAkeN

 

oN

 

oldeR

 

cARs

 

WITh

 

NoN

-

GAlvANIzed

 

sTeel

 

bodyWoRk

The

 „s

Teel

” 

opTIoN

 

shAll

 

be

 

choseN

Zeroing (calibration)

Before You start to take measurements, please check whether the gauge is calibrated. To do it 
properly, please select the 

Zero

 function from the main menu and put the sensor against the 

green side of the calibration sheet. If the display shows a 0 +/-10µm value, it means that the 
gauge is calibrated. Should the aberration be larger than 0 +/-10µm, please put the sensor once 

Summary of Contents for TW-6

Page 1: ...Thoma Tel Fax 48 95 741 11 62 info t4w eu www t4w eu Coating Thickness Gauge TW 6 Operation Manual a r t n r 5 9 2 4 8...

Page 2: ...peration Manual TW 6 Table of contents 1 Technical specifications 3 2 Getting started 3 3 Gauge manual 4 Guarantee TW 6 9 The price of this product includes the recycling fee of 0 10 PLN Before You st...

Page 3: ...ent is made The memory doesn t get deleted after turning off the device although it is possible to delete it from the menu The device has a LCD backlight which makes measuring in places with insuffici...

Page 4: ...nstead of a alkaline battery You can use a compatible fully charged accu 3 If the device seems to work improperly it can be caused by an empty battery or accu To take a measurement please put the devi...

Page 5: ...ion function switches between the measurement resolution Soft indicates the installed software version To select a function please press the OK MENU button Measure function Measur After You select thi...

Page 6: ...ed You have the following possible choices galvanized steel aluminium Zn Al steel Fe reference point Ref You can switch between the sheet materials with the FUNKTION button to accept the selection pre...

Page 7: ...means that the gauge is well calibrated Should the aberration be larger than 0 10 m please put the sensor once more against of the calibration sheet wait until the result stabilizes and press the OK M...

Page 8: ...choose between 2 measurement resolutions chosen with the FUNKTION button D 10 the measurement is rounded up to 10 m faster D 1 the measurement is accurate up to 1 m precise For car paint coating thic...

Page 9: ...be shipped to the Producer it happens on the shipper s responsibility and cost 8 The device will be not accepted by the Service if pt 6 is not kept if the defect is of no workmanship or 9 material nat...

Reviews: