background image

 

the  manual  range  function, if  the  measurement   range  is  unknowed ,  press 
RANGE   key to the maximum range, first  press the INRUSH button 2 seconds again 
inrush measurement will be resumed, and the "INR" symbol disappear. 

ZERO:zero key of DC current function, when  measuring DC current and it is not 

display  zero  without  measuring,press  ZERO  key  to    make  it  display  zero  before 
measure. 

DH: Data  hold button: The user  may hold the present reading and keep it on the 

display  by  pressing  the  “DH”  button.  Press  the  button  again  to  concel  the  data-hold 
function 

7

LCD display 

8

“V

”jack:  This  is  positive  input  terminal  for  voltage,resistance 

frequency,temperature,capacity and diode. 

9

 “COM”jack:This is negative(ground)input terminal   

(10)    Barrier 

 

4.2

 Measurement of AC/DC voltage

 

Turn the Rotary  switch to “V”. Then plug black lead in “COM” socket, and plug 

red  lead in  “V/

”  socket.  connect  the  test  lead  with  the  two  ends  of  the  circuit  and 

then directly read the reading on the LCD display   
NOTE: 1. The inrush  measuring funciton can only be change range  manually, before 
the test,if the voltage is unknown, manually set the measuring range to 600V and then 
press the PH key to enter the INRUSH measuring function.   
2.  Don't measure  the  peak voltage more  than  600V, otherwise  it  might  damage  the 
instrument ,if the  screen only displays  OL, it  means that the tested voltage is  higher 
than 600V. 
3.  Presse  the  SRLECT  key  in  the  ACV  function  to  enter  the  voltage  frequency 
measurement  mode,  it  can  measure  the  the  frequency  range  from  10Hz  to  100KHz 
with voltage higher than AC 30V

~

600V

4.3.1  DC current 

measurement

 

Turn rotary switch to

DCA

range, if it isn’t displaying zero ,press 

ZERO

button until zero displayed on the LCD.   

Press the trigger, opens the  mouth of the clamp, and  hold a  wire ( put the  wire  in 

the clamp center ), read the reading directly. 

4.3.2

 

AC current measurement 

Turn  rotary  switch  to

~A

  range.  Press  the  trigger,  opens  the  mouth  of  the 

clamp, and hold wire ( put the wire in the clamp center ), read the reading directly 
        Note:1.INRUSH  measuring   can  only  be  manually setting range.  when  the 
current value being measuring is unknow, turn    the rotary switch to 600A    and press 
INRUSH    key to entry    INRUSH measuring funciton. 
        2.  When  measuring  current  ,the  clamp  might  hold  only  one  wire,it  is  null  to 
catch one more another wire 

4.4

 

Measurement of resistance,continuity and forward Voltage drop of diode

 

Warning!  When  measurement  of  resistance  or  continuity,  make  sure  that  no 

voltage is with the circuit and components.

 

1

Turn  the  Rotary  switch  to  the  range  of 

Ω

/

/

.  At  this  time,  the  meter  is 

reserved at resistance range. 

2

Plug red lead in “V/

Ω

” socket, and plug black lead in “COM” socket 

3

Connect the leads with the two ends of the circuit or component, and then read 

the value of resistance.   

4

Press SELECT key to change the range of 

. When the resistance measured 

is less than about 90

±20

, the buzzer sounds. This is continuity testing.   

5

When the test lead is under open-circuit or input-overload status, the display will 

display “OL”. 

6

When measuring diode,press the SELECT key to the function of 

.   

7

Connect the test lead with the two ends of the diode, and then read the value of 

forward voltage drop value.   

8

When the test lead is under reverse connection or open-circuit status, the display 

will display “OL”. 

  

Note

a

When the resistance measured is above 1M

Ω

, it needs several seconds for 

the reading to be stable, it is normal when measuring high resistance. 
b

When measuring high resistance, insert the resistance pin directly into the V Ω and 

COM jack, so as to avoid interference 
c

When measuring resistance in a circuit make sure the power to the circuit is turned 

off and all capcitors are discharged. 

4.5  

Measurement of capacitance

 

 

Warning!  When  measurement  of  capacity,  the  measured  capacitor  should  be 

completely discharged. 

 

(1)

Turn the Rotary  switch to “

” function. Plug red lead in “V/

Ω

” socket, and 

plug black lead in “COM” socket.

 

Warning:The  range  for  capacitor  can’t  be  set  manually.  When  the  capacity  value 

is large, the time for measurement may be a little longer.

 

a. 

 

  Do not take an external voltage or charged capacitor (especially a large 

capacitor) connected to the test terminal   

b.    When a large capacitor is serious leakage or breakdown, the measurement value 

may be instability 

4.6

 Measurement of frequency

 / 

DUTY Ratio

 

1

Turn  the  Rotary  switch  to  “Hz”  function.if  you  want  to  measure  DUTY   

Ratio,Press SELECT key to switch   

2

Plug red test lead in “V/

Ω

” terminal and plug black lead in “COM” terminal. 

3

Connect the leads with measured circuit and then read the reading. 

4

when  the  voltage  exceeding  30V,  please  enter  the  Voltage   frequency 

measurement   mode  by  pressing  the  SELECT  key  in  the  “ACV”  measurement 
function. 

4.7 NON    CONTACT    VOLTAGE      TESTING

           

Set the rotary switch at the desired “ ” range position.NCV and ~symbol will be 

 displayed on the LCD,connect the red test lead to “VΩ” jack and the black test lead 
without being connected ,hold the red test lead to approach the phase line 
of   commercial power ,switch,or charge,(hold the red test lead  to touch the metal ter
minal if it is necessary),the meter will display    “ ”    ,the higher the tested voltage is , 
the more " " will be displayed,and the buzzer will warning hurrier,       If the red test 
lead touch the conducter, when there is voltage between null line and ground line, 
much more " "will be displayed when measuring the phase line than the null line. 

When the meter has no response to the voltage,you can contect the red test lead to

 the metal of the terminal. 

 

Caution:   

1.even if there is no indication, voltage may still exist. Do not judge the wire whether 
threre is voltage absolutely throught the non contact voltage testing,the testing may be 
effected by many factors such as the socket design,the insulation thickness and types 
etc. 
2.Interference source of external environment,such as flash,motor etc,may false 
triggle the non contect voltage testing. 

 

. Maintenance of meter 

 

 

Warning!  Switch  off  the  power,  remove  the  test  lead  and  any  of  input 

signals to prevent electric shock before opening the cover of meter or the cover of 
battery. 
 
5.1    When the  meter displays the  symbol of “

”, the battery should be changed. 

Open the battery cover, and then change the used  battery  with  new  battery to ensure 
the normal operation of the meter.   
5.2    Keep  the  meter  and  the  pens  clean,  dry  and  non-destructive.  Clean  cloth  or 
detergent  may  be  used  for  cleaning  the  cover  of  the  meter.  No  grinding  agent  or 
organic solvent can be used for the same cleaning purpose.   
5.3    The  meter  should  be  protected  against  damage,  vibration  and  impact.  It 
shouldn’t be placed where high temperature or intense magnetic field exists.   
5.4 

Calibrating of the meter is done on a yearly basis.     

 

Accessories   

6.1    Test lead: 1 set   
6.2    Users manual: 1 piece 
6.3    Cloth bag: 1piece 

 