Technical specifications
EUROLYZER® ST / STe
13
Parameter
Value
Sensor
Thermocouple NiCr-Ni (type K)
Pressure measurement
Measuring range
± 50 hPa (draft)/± 130 hPa (pressure)
Max. deviation
± 2 Pa + 1 digit (0 hPa to ± 2.00 hPa)
Resolution
± 1 % of measured value (± 2.01 hPa to
± 50.0 hPa)
± 1.5 % of measured value (± 50.1 hPa to
± 130.0 hPa)
Sensor
Semiconductor sensor
O
2
measurement
Measuring range
0-21.0 % by volume
Max. deviation
± 0.2 % by volume of measured value
Resolution
0.1 % by volume
Sensor
Electrochemical measuring cell
Response time (T90) 50 seconds
CO
2
determination
Range
0 to CO
2 max
(fuel-specific)
Max. deviation
± 0.2 % by volume of measured value
Resolution
0.1 % by volume
Sensor
Calculation on the basis of measured O2
value
Response time (T90) 50 seconds
CO measurement (with H
2
compensation)
Measuring range
0-5000 ppm (nominal)
or 9999 ppm (maximum)
Accuracy
5 ppm (to 50 ppm)
5 % of measured value (above 50 ppm)
Resolution
1 ppm
Sensor
Electrochemical measuring cell
Response time (T90) 60 seconds