System Sensor FAAST FL0111E-HS Quick Installation Manual Download Page 9

D200-101-01

9

I56-6574-000

Turn the sensor head anti-clockwise to remove from base (and 

clockwise to relocate). Do 

NOT

 interchange the devices and do 

NOT

 

alter the rotary address switch settings on the sensors.  If replacing a 

sensor, ensure that the address set on the new sensor is the same as 

on the sensor being replaced. Use 

ONLY

 model number F-SEN-SSE 

as a replacement sensor.

WARNING

Using compressed air to clean the pipe system

High pressure air flushed through the system could damage the 

fan, ensure that the FAAST LT-200 unit is sealed or detached from 

the system before commencing this procedure.

 

USB Connection

PC connectivity is provided by an on board USB 

B

 port located centrally 

between  the  filter  and  the  sensor  cover  (see  Figure  9  displayed 

earlier in the guide). The USB interface allows access to the PipeIQ 

application software running on a PC.

Note:  The  USB  connecting  cable  should  be  removed  during 

normal operation

.

Changing Default Settings / Verifying a Pipe Layout Design

To change any of the default options, or to use the pipe layout design 

function, it will be necessary to connect the detector to a PC/laptop 

with the PipeIQ software installed (version 2.9.1 or higher). For more 

complex designs using increased functionality (experienced users 

only) see the 

FAAST LT-200 Advanced Setup and Control Guide

 for 

further information.

P

ipe

IQ

 SOFTWARE

The PipeIQ software is supplied on a USB pen stick with the device, 

or can be downloaded from www.faast-detection.com. 

For the FAAST 

LT-200, it is necessary to use PipeIQ version 2.9.1 or higher.

Note 1:

 It is strongly recommended that some form of training is 

taken before using any version of PipeIQ (contact your distributor 

for information on this)

. The information provided here is only a brief 

introduction.

Note 2:

 To connect a FAAST LT-200 to a PC using the USB port, the 

PipeIQ software must be running on the PC and the device must be in 

Maintenance

 mode (See 

Password Sequence to Enter Maintenance 

Mode

 section previously).

Connecting a PC to a FAAST LT-200 Device for the First Time

Minimum System Requirements

Microsoft Windows Vista, 7, 8 or 10, XP - SP3 (not recommended). 

1 GB of RAM.

Graphics hardware with 128 MB of memory and support for OpenGL 

2.0 or later.
5 GB of free hard disc space

When the PipeIQ installation is complete:

1)  With PipeIQ open on the PC and the device in 

Maintenance

 mode, 

open the device front door by releasing the two Phillips screws. 

2)  Connect a USB cable from the internal port in the centre of the 

device to a spare USB port on the PC (make sure that the USB 

connector is pushed fully into the FAAST LT-200 port until it clicks). 

The device will now be in 

Remote Maintenance

 mode.

  The USB connection must be made within 5 minutes of entering 

the 

Maintenance

 mode password.

 

The first time a PC running PipeIQ is connected to a FAAST LT-200 

device, the USB port drivers will be installed.

3)  In PipeIQ, open the project file (.mdf file) that is associated with 

the FAAST LT-200 device being used. If no project file has been 

created yet, click on 

New

 and create a new project. Accept the 

PipeIQ disclaimer, select the measurement units, device type and 

number of channels. 

Make sure that the device type selected 

matches the type of FAAST LT-200 device being used

. The 

project will assume the factory default configuration for a device.

4)  In the Left Hand pane of PipeIQ, left click on the FAAST LT-200 

Device in the system tree to highlight it. Right click on the FAAST 

LT-200 icon in the left hand pane to reveal the menu. Click on 

Connect Device

  Note that to use the full range of connection options, a user must 

be connected to the FAAST LT-200 device as an 

Administrator

Tick the 

Admin

 box and enter the 

Administrator

 password to 

access these extra commands. The 

Administrator

 password is the 

same as the 

Remote Maintenance

 password (the default is 3111).

5)  Click on 

Connect

The detector is connected when a small green 

tick is indicated on the device icon. 

FAAST LT-200 Connected

 is 

shown at the bottom left corner of the screen.

6) Depending on what operations are required, select the 

Configuration

Design

 or 

Monitoring

 tab at the bottom of the screen 

as necessary.

For greater detail on using PipeIQ with the FAAST LT-200 see the 

FAAST LT-200 Setup and Troubleshooting Guide

. For information on 

using the 

Configuration

Design

 and 

Monitoring

 tabs in PipeIQ, see the 

FAAST LT-200 Advanced Setup and Control Guide.

Setting the Fan Speed

A default air velocity reference value is set into each FAAST LT-200 

unit before leaving the factory. This equates to an optimum air flow of 

approximately 45 l/min. Default high/low flow limit thresholds are set to 

guarantee a flow fault when the air flow is ±20% of the reference flow.

The FAAST LT-200 fan speed in each channel can be set to 

Auto

 or 

Manual 

control. Use Auto mode in normal operation.

EN54-20 : 2006

Class A B & C

Aspirating Smoke Detectors

0832  18

FL0111E-HS:  DOP-ASP022

FL0112E-HS:  DOP-ASP023

FL0122E-HS:  DOP-ASP024

System Sensor Europe

Life Safety Distribution GmbH

Javastrasse 2, 8604 Hegnau, 

Switzerland

DoP Ref:

System Sensor Europe

Pittway Tecnologica S.r.l. 

Via Caboto 19/3

34147 TRIESTE

Italy

Summary of Contents for FAAST FL0111E-HS

Page 1: ...it Warning The performance of this system is dependent upon the pipe network Any extension or modification to the designed installation may cause improper operation The operational effects of such changes need to be verified using the PipeIQ design software This equipment and all associated pipe work must be installed in accordance with all relevant codes and regulations PARTS LIST Description Qua...

Page 2: ... place remove the protector from the bottom of the clear cover to stick the cover down as shown in Figure 3 Figure 5 Mounting Bracket FAAST LT 200 HANGING LUGS INDICATES CENTRE OF ASPIRATING PIPES FIXING HOLE FIXING HOLE USE FOR PLUMB LINE 90 mm 99 mm 59 329 212 mm 144 mm 41 mm mm mm POSITION OF BRACKET PLUMB LINE BASE OUTLINE OUTER DIMENSION OF UNIT 1 2 3 4 5 6 7 8 9 10 1 11213 Minimum clearance ...

Page 3: ...es not used should be kept sealed Note 2 Do NOT glue pipes into the pipe holes Table 1a Maximum Number of Pipe Holes Allowed Per Channel All figures quoted using highest level 1 sensitivity 1 2 3 4 5 6 CHANNEL 1 CHANNEL 2 ONLY FUNCTION ON 2 CHANNEL UNITS CHANNEL 1 NOT USED FOR COMMON CHAMBER UNIT CHANNEL 2 ONLY FUNCTIONS ON 2 CHANNEL AND COMMON CHAMBER UNITS FAAST LT MODEL INLET PIPE HOLE OUTLET P...

Page 4: ...f the unit to lever open the door as shown above SAMPLING AREA SAMPLING AREA Whenever the FAAST LT 200 is installed outside the risk area return of the exhaust air back into the protected area can reduce flow faults due to pressure difference MODULE MOUNT MODULE MOUNT F LT PMB MODULE MOUNT KIT OPTIONAL If required an input output module can be installed into the FAAST LT 200 unit The optional modu...

Page 5: ...CIFICATION MIN MAX UNITS COMMENTS Contact Rating 2 0 5 A A 30 VDC resistive load 30 VAC resistive load Life Time 105 Operations 1N6284CA FAAST LT FAAST LT C L FAAST LT R C L WARNING Switching Inductive Loads Fitting the Terminal Blocks To insert the terminal blocks into the unit use the following method 1 Insert a corner of the block into the slot see a 2 Push the length of the block into the slot...

Page 6: ...A TIO N DIS AB LE SY ST EM S O U N D E R F IL T E R L O W F L O W H IG H F L O W INITIALIZATION LEVEL SMOKE FAULT FAULT POWER ALARM PREALARM 2 10 3 1 4 5 6 7 8 9 TEMPERA TURE INPUT 10b FL0122E HS 2 Channel Detector POWERING UP Using Default Settings 1 Connect a suitable 24VDC supply complying with European Standard EN 54 4 to pins 1 and 2 on terminal block T1 See Table 2 2 Check the voltage at the...

Page 7: ... INITIALIZATION ALARM PREALARM 2 10 3 1 4 5 6 7 8 9 2 10 3 1 4 5 6 7 8 9 INDICATOR ACTION WARNING OR TROUBLE COMMENT ACTION CHANNEL 1 2 ALARM ON Red Channel is in alarm relay is set No delay with default settings 1 BLINK Green when sensor is polled Not in alarm CHANNEL 1 2 PRE ALARM ON Yellow Channel is in pre alarm relay is set SMOKE LEVEL ON Yellow Led number indicates sensor alarm level reached...

Page 8: ...troduce smoke particulates into the system It is recommended that smoke with a particulate life cycle of greater than 120s should be used standard aerosol sprays for point detector testing do not work well on aspirated systems Fault Testing Simulate a fault on the detector for example block the outlet pipe and check that a fault is signalled on both the front panel of the unit and at the CIE Fire ...

Page 9: ...ernal port in the centre of the device to a spare USB port on the PC make sure that the USB connector is pushed fully into the FAAST LT 200 port until it clicks The device will now be in Remote Maintenance mode The USB connection must be made within 5 minutes of entering the Maintenance mode password The first time a PC running PipeIQ is connected to a FAAST LT 200 device the USB port drivers will...

Page 10: ... 45l min 6 Repeat from step 3 adjust hole diameters and fan speed to achieve all the above figures 7 Next check that the sample hole sensitivity is practical for the chosen class i e it is not too sensitive Ideally to avoid false alarms hole sensitivity for a class C system should be 1 m or greater and certainly 0 5 m 8 If necessary change the Alarm level to reduce the sample hole sensitivity The ...

Page 11: ...76 98 60 6 4 4 45 0 6 1 14 3 10 87 98 50 5 4 5 4 5 45 0 62 0 98 3 9 95 44 40 4 4 5 4 5 45 0 76 0 80 3 10 79 53 30 3 5 5 46 0 83 1 29 4 10 78 55 20 2 5 5 5 5 45 0 92 0 97 4 10 71 28 10 1 6 5 6 5 45 1 1 31 5 10 66 37 Non sensing end hole CLASS C PIPE DESIGNS WITH SINGLE SAMPLE HOLE END HOLE SIZE AND 10m SPACING 2 PIPE T FORM Per pipe length Number of holes Hole size End hole Air flow Flow balance Av...

Page 12: ...pe Number of holes Hole size End hole Air flow Flow balance Average sensitivity Alarm level Fan Total hole area m per pipe mm mm I min m mm2 50 3 5 5 45 0 84 0 40 1 4 117 83 40 3 4 5 4 5 46 0 89 0 40 1 5 95 44 30 3 4 4 46 0 93 0 40 1 7 75 41 20 2 5 5 46 0 96 0 40 2 6 78 55 10 1 6 5 6 5 44 1 0 33 3 8 66 37 CLASS B PIPE DESIGNS WITH SINGLE SAMPLE HOLE SIZE Inc SAMPLING END HOLE AND 10m SPACING 1 PIP...

Page 13: ... 83 0 20 1 9 84 83 20 2 6 5 6 5 44 0 94 0 14 1 10 66 37 10 X X No adequate or recommended single sample hole size design CLASS A PIPE DESIGNS WITH SINGLE SAMPLE HOLE SIZE Inc SAMPLING END HOLE AND 10m SPACING 2 PIPE T FORM Each pipe Number of holes Hole size End hole Air flow Flow balance Average sensitivity Alarm level Fan Total hole area m per pipe mm mm I min m mm2 50 3 5 5 45 0 84 0 40 1 4 117...

Reviews: