21
Directions For Use
Appendix A Immunity All Equipment and Systems
Immunity Test
EN/IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment – Guidance
ESD
±6kV Contact
±6kV Contact
Floors should be wood,
EN/IEC 61000-4-2
±8kV Air
±8kV Air
concrete or ceramic tile. If floors are
synthetic, the r/h should be at least 30%
EFT EN/IEC 61000-4-4
±2kV Mains ±1kV
I/Os
±1kV Mains ±1kV
I/Os (see
justification)
Surge
±1kV Differential
±1kV Differential
Mains power quality should be that of a
EN/IEC 61000-4-5
±2kV Common
a typical commercial or hospital environment.
Voltage Dips/Dropout
>95% Dip for
>95% Dip for
Mains power quality should be that of a
EN/IEC 61000-4-11
0.5 Cycle
0.5 Cycle
typical commercial or hospital environment.
60% Dip for
60% Dip for
If the user of the v900L VTS-1000 charging
5 Cycles
5 Cycles
charging system requires continued
30% Dip for
30% Dip for
operation during power mains interruptions,
25 Cycles
25 Cycles
it is recommended that the charger be
>95% Dip for
>95% Dip for
powered from an uninterruptible power
5 Seconds
5 Seconds
supply or battery.
Power Frequency
3A/m
3A/m
Power frequency magnetic fields should
50/60Hz
be that of a typical commercial or
Magnetic Field
hospital environment.
EN/IEC 61000-4-8
Emissions Equipment and Systems that are NOT Life-supporting.
Immunity Test
EN/IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment – Guidance
Conducted RF
3 Vrms
(V1)Vrms
EN/IEC 61000-4-6
150 kHz to 80
MHz
Portable and mobile communications equipment
should be separated from the v900L by no less than
the distances calculated/listed below:
D=(3.5/V1)(Sqrt P)
Radiated RF
3 V/m
(E1)V/m
D=(3.5/E1)(Sqrt P)
EN/IEC 61000-4-3
80 MHz to 2.5
GHz
80 to 800 MHz
D=(7/E1)(Sqrt P) 800 MHz to 2.5 GHz where P is
the max power in watts and D is the recommended
separation distance in meters. Field strengths from
fixed transmitters, as determined by an
electromagnetic site survey, should be less than the
compliance levels (V1 and E1). Interference may
occur in the vicinity of equipment containing a
transmitter.
Max Output Power (Watts)
Separation (m)
150kHz to 80MHz
D=(3.5/V1)(Sqrt P)
Separation (m) 80
to 800MHz
D=(3.5/E1)(Sqrt P)
Separation (m)
800MHz to
2.5GHz
D=(7/E1)(Sqrt
P)
0.01
.1166
.1166
.2333
0.1
.3689
.3689
.7378
1
1.1666
1.1666
2.3333
10
3.6893
3.6893
7.3786
100
11.6666
11.6666
23.3333
Recommended Separation Distances between portable and mobile RF Communications
Equipment and the v900L Equipment and Systems that are NOT Life-supporting.
v900L