Synthes
13
7
Reaming guidelines (optional)
Instruments
351.706S* 2.5 mm Reaming Rod with ball tip,
950 mm, sterile
360.255 Reaming Rod Measuring Device
Using image intensification, ensure that fracture reduction
has been maintained. Insert the 2.5 mm reaming rod
with ball tip into the medullary canal to the desired
insertion depth.
Ream in 0.5 mm increments and advance the reamer
with steady, moderate pressure. Do not force the reamer.
Partially retract the reamer often to clear debris from the
medullary canal.
Ream to a diameter at least 1.0 mm greater than the nail
diameter as determined by surgeon preference.
After reaming, remove the reaming assembly, leaving the
reaming rod in place.
Note:
The trochanteric fixation nail can be passed over
the 3.0 mm reaming rod, with straight ball tip, if used.
No reaming rod exchange is required.
Nail length may be determined by using the reaming rod
measuring device and a 950 mm reaming rod or guide rod.
Insert the reaming rod to hold fracture reduction. Position
the image intensifier over the distal femur and take an image
to confirm reaming rod insertion depth. Pass the reaming
rod measuring device over the proximal end of the reaming
rod and through the incision to the bone. Read nail length
directly from the measuring device.
* Also available