Troubleshooting Guide – Physician’s actions:
Error Action
Low Battery
Replace the control system with a new one.
System error
Replace the control system with a new one.
End of treatment
Please verify that the struts are at their final position according to plan.
Remove the control system from the patient. Attach strut gear lockers to each strut.
See section 4.1.4.
Treatment does not
Replace the control system with a new one. If a replacement is not available, remove
occur (for example: struts
MAXFRAME AUTOSTRUT Control System and Struts and replace with MAXFRAME
do not move/dead system, Quick Adjust or Standard Struts. The MAXFRAME AUTOSTRUT System is backwards
hardware failure, or unable compatible with the manual struts found within MAXFRAME Multi-Axial Correction
to upload treatment plan.)
System. Refer to the MAXFRAME surgical technique guide for details.