CHAPTER SEVEN
7-2
bc620/627AT Time and Frequency Processor
Symmetricom, Inc.
7.2 TIME CODE OUTPUT AMPLITUDE ADJUSTMENT
RV1 adjusts the amplitude of the modulated IRIG B output time code. A value of one volt RMS
is common as is three volts peak-to-peak on the high cycles. Adjust this value to suit the
equipment being driven. The range is zero to twenty-four volts peak-to-peak.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com