Installing and Setting Up the Symmetricom 5087B
2
Symmetricom 5087B Operations and Maintenance Manual Rev. E
19
The hardware device number and IP address appear in the
Assign IP
Address
dialog box.
6
Type the new IP address and click
OK.
The new IP address appears in the Lantronix® DeviceInstaller
window.
7
Test the IP address by pinging the instrument’s Xport™ on the
Lantronix Xport Installer window.
Click the
Ping
icon or select
Ping Device
on the
Tools
menu.
The
Ping Device
window should show the IP address of the
instrument’s Xport™ device, and it should show successful replies if
the IP address has been configured correctly.
8
Exit the browser.
To assign the static IP address using Telnet
1
Telnet to the assigned address, port 9999.
2
Press
Enter
within five seconds to enter the setup mode.
3
Select Option 0.
4
Set the IP address and follow the on-screen instructions to save the
setting.
5
Telnet to the new IP address, port 10001.
6
Type
I
You do not need to type a carriage return or line feed.
The system returns
I5087B-00\r\n
, where the 00 is the hardware
version, and turns on the front panel input LEDs for two seconds
before returning to normal operation.
If you see these responses, you know you are communicating with
the unit.